MAGSTIM 200 2 Operating Manual page 18

Table of Contents

Advertisement

2
Magstim 200
Error codes produced by base system.
Error Codes
E61
E62
E63
E64
E65
E66
E67
E68
E70
E71
E72
E73
E74
E75
E76
E77
E78
E79
E80
E81
E82
E83
E84
E85
E86
E87
E88
E89
E90
The Magstim Company Limited 3001-23-04
Stack underflow
Unexpected reset
SYS debug error
Coil under temperature
Coil max difference
Stack over temperature
Stack under temperature
Stack max difference
HVCAP over temperature
HV Transformer over temperature
Charge threshold failure
VREF check failure
HVCAP voltage comparison failure
HV over voltage
Invalid system configuration
Base coil stop line fault
UI stop line fault
Dump system fault
WARNING – high voltage may be present up to 20
minutes after use on internal high voltage capacitor.
Invalid coil category for current system configuration.
Invalid NVRAM Checksum.
Faulty charging relay detected
Base arm LED drive failure
16
Description
Power fail
Stack overflow
RTS overrun
WD Timeout
Bad checksum
Charging fault
Stop line fault
LUMENIS EX1091