MAGSTIM 200 2 Operating Manual page 9

Table of Contents

Advertisement

2
Magstim 200
REPLACE COIL Indicator (9)
DEFAULT CONDITION Indicator (10)
ARMED Indicator (11)
READY Indicator (12)
The Magstim Company Limited 3001-23-04
This red LED will illuminate if any of the following conditions are
present:
1. The Coil connected is not compatible with the Magstim 200
2. The surface temperature of the connected Coil has risen above its set
limit.
3. The connected Coil is faulty.
4. There is no Coil connected.
If this LED is illuminated the Magstim 200
inactive default condition.
This LED is illuminated when the Magstim 200
safe inactive default condition.
This LED is illuminated when the unit is in an armed condition. This
indicates that there is high voltage present inside the unit.
This LED is illuminated when the Magstim 200
be triggered. A magnetic pulse can be produced by depressing one of the
Trigger Enable switches at the base of the stimulating coil and
depressing the TRIGGER button (15), or by depressing the footswitch,
or by applying a Trigger Input to the TRIGGER INPUT connector on
the rear panel.
Note: If the instrument has not been triggered for over 1 minute after the
READY symbol is illuminated the unit will automatically select the
default mode and discharge internally.