2
Magstim 200
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
2
The Magstim 200
is intended for use in the electromagnetic environment specified below. The
customer or the user of the equipment should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Electrostatic
6 kV contact
discharge (ESD)
8 kV air
IEC 61000-4-4
Electrical fast
2 kV for power
Transient/ burst
supply lines
IEC 61000-4-11
1 kV for input/
output lines
Surge
1 kV differential
mode
IEC 61000-4-5
2 kV common
mode
Voltage dips,
<5 % U
short interruptions
(>95 % dip in U
and voltage
0,5 cycle
variations on
power supply
40 % U
input lines.
(60 % dip in U
5 cycles
IEC 61000-4-11
70 % U
(30 % dip in U
25 cycles
<5 % U
(>95 % dip in U
5 sec
Power Frequency
3 A/m 50Hz
(50/60Hz)
Magnetic Field
IEC 61000-4-8
Radiated RF
80MHz – 2.5GHz
Immunity
2Hz 80% amplitude
EN61000-4-3
modulation
Conducted RF
0.15MHz – 80MHz
Immunity
2Hz 80% amplitude
EN61000-4-6
modulation
NOTE U
is the a.c. mains voltage prior to application of the test level
T
The Magstim Company Limited 3001-23-04
Compliance
Level
Meets requirement
Meets requirement
Meets requirement
Meets requirement
T
) for
T
T
) for
T
T
) for
T
T
) for
T
Meets requirement
Meets requirement
Meets requirement
Electromagnetic environment -
guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Magstim
2
200
requires continued operation during
power interruptions, it is recommended that
2
the Magstim 200
be powered from an
interruptible power supply.
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment
Equipment should only be used in the
vicinity of other equipment compliant with
EN60601-1-2.
Equipment should only be used in the
vicinity of other equipment compliant with
EN60601-1-2.
31
LUMENIS EX1091
Page 33
th
18
March 2005