MAGSTIM 200 2 Operating Manual page 13

Table of Contents

Advertisement

2
Magstim 200
TRIGGER INPUT/ OUTPUT
FOOT SWITCH Pneumatic Socket (3)
COOLING FAN OUTLET (4)
POWER ENTRY MODULE (5)
FUSE ACCESS & VOLTAGE SELECTOR (6)
The Magstim Company Limited 3001-23-04
The trigger input is used to synchronise the Magstim 200
trigger input. Application of a 5V CMOS Logic Level to this input will
result in the instrument being triggered, provided the READY symbol
(12) is illuminated and one of the safety buttons at the base of the
stimulating coil is activated. This signal must be active for at least 10µs.
In most instances the Magstim 200
recording instrument supplied by various manufacturers. In these cases
the Magstim and the recording instrument will have to be synchronised so
that the magnetic pulse produced by the Magstim occurs exactly at the
start of the recording sweep of the recording instrument. Almost all
recording instruments have some means of accepting an external trigger
signal via either a dedicated input port or via the foot switch trigger input.
The trigger output supplies a synchronisation signal to the external
recording instrument. A 100µs 5V CMOS Logic Level pulse of either
polarity is provided at this output whenever the Magstim is discharged.
Attention: Consult accompanying documentation before using the
2
Magstim 200
.
This socket is used to connect the supplied pneumatic FOOT SWITCH
2
. The FOOT SWITCH can be used to trigger the
to the Magstim 200
instrument when one of the safety switches at the base of the stimulating
coil is activated and the READY symbol on the front panel is
illuminated.
With the Magstim 200
of the instrument and expelled through the FAN OUTLET. Since it is
important to allow unrestricted air flow, the instrument must be placed on
a hard surface. In addition, there should be no obstructions placed closer
than 50mm to the FAN OUTLET.
See Section 9 of this operating manual for instructions on how to replace
the fuses. NB The fuses should only be replaced with those specified in
this operating manual.
2
will be used in conjunction with a
2
turned on, air is drawn through slots at the bottom
11
2
to an external
LUMENIS EX1091
Page 13
th
18
March 2005