MAGSTIM 200 2 Operating Manual page 8

Table of Contents

Advertisement

2
Magstim 200
ARMED Indicator (2)
Unit Power Status Indicator (3)
OUTPUT % DISPLAY Indicator (4)
COIL IN Indicator (5)
COIL TEMPERATURE Indicators (6, 7 & 8)
The Magstim Company Limited 3001-23-04
This LED is illuminated when the unit is in an armed condition. This
indicates that there is high voltage present inside the unit. On power-up
the armed indicator on the main unit flashes three times as part of a self-
test programme.
Condition 1. – While the Magstim 200
controlled by the Mains Power switch (1), this LED flashes at
approximately 0.5Hz, indicating that the unit is in a standby condition.
Condition 2. – While the Magstim 200
controlled by ON/ OFF (Standby) button (2), this LED is continuously lit,
indicating that the unit is in the Operational state.
This display is used to show the following:
1. The set power output as a percentage of maximum. After turning the
unit on using the ON/OFF button (1) the OUTPUT % DISPLAY will
display 30%. The OUTPUT CONTROL knob (16) can then be used to
change the output level in 1% steps. Note: The OUTPUT % DISPLAY
only shows the level the Magstim 200
READY symbol (12) is illuminated.
2. The system status codes. These are detailed in Section 6 – System
Status Codes.
This green LED will illuminate if there is a coil detected as being
connected correctly to the Magstim 200
These LED's represent the surface temperature of the coil. The LED's
illuminate sequentially as the coil temperature rises, beginning with the
bottom green LED and ending with the amber LED. When the amber
LED is lit, it means that the coil is getting close to its maximum operating
temperature. As soon as it reaches its maximum operating temperature,
the replace coil indicator will illuminate and the M200
a safe default mode and discharge internally.
NB If the temperature of the coil is below 5
operate. The Magstim 200
until it detects that the coil temperature has risen above 5
2
is in the Standby state, as
2
is in the Operational state, as
2
is actually charged to if the
2
.
o
C, the system will not
2
will remain in its inactive default condition
6
2
will put itself into
o
C.
LUMENIS EX1091
Page 8
th
18
March 2005