MAGSTIM 200 2 Operating Manual page 16

Table of Contents

Advertisement

2
Magstim 200
8. Press the RUN button. The ARMED indicator will illuminate and when
the instrument is charged to 30% the READY indicator will also illuminate.
This indicates that the instrument is ready for use.
Note: If the instrument has not been triggered for over 1 minute after the
READY symbol is illuminated the unit will automatically select the default
mode and discharge internally.
9. Hold the stimulating coil and press one of the Trigger Enable switches
located on either side of the coil handle. Press and hold the TRIGGER
key on the front panel, or depress and hold the footswitch. There should be
a clicking noise emanating from both the stimulating coil and the Magstim,
each time the Magstim is discharged, indicating that a magnetic pulse is
being produced.
Note: When using a standard 2
the safety switches at the base of the stimulating coil will enable the armed
M200
item 15 in Section 3 (front panel description) of this operating manual, or
the footswitch – item 3 in Section 4 (rear panel description) of this
operating manual. If a Remote Control Coil is being used, depression of
one of the safety switches will cause the system to arm, depression of the
second switch will trigger the system.
On no account must the trigger enable buttons on the stimulating coil
be permanently activated, by taping down, or any other method of
fixing. They are designed as a safety interlock, to disable them could
result in a serious injury as a result of an inadvertent discharge of the
coil.
Ensure the instrument is functioning properly before proceeding.
Note: The Magstim 200
TRIGGER INPUT Port, located on the rear panel of the unit.
Completion of Session:
During prolonged use, the temperature of some of the internal components
of the Magstim 200
the operation of the cooling fan. However, when the unit is switched off,
the cooling operation is terminated. Therefore, it is recommended that, if
the Magstim 200
switched on for approximately ten minutes after the last session has been
completed, to allow the fan to cool the unit.
The Magstim Company Limited 3001-23-04
2
, allowing it to be triggered either via the trigger button on the unit –
2
can also be triggered by applying a signal to the
2
will tend to rise. During operation this is regulated by
2
has been subject to prolonged use, the unit is left
14
nd
Generation Coil, pressing one, or both, of
LUMENIS EX1091
Page 16
th
18
March 2005