Operating Design
Parameter
Settings
ON Z/T
ON*
OFF
ISOCAL
ON*
OFF
NOTE
CAL.UNIT
GRAMS*
CONF.UNIT
* Factory setting
4.8.2

Parameters in the "SETUP/GEN.SERV." Menu

Parameter
Settings
MEN.RESET
DEFAULT
NO*
* Factory setting
32
Operating Instructions Cubis
Explanation
Activates the initial taring | zeroing. The device is tared or zeroed after it is
switched on.
Deactivates the initial taring | zeroing: After it is switched on, the device shows
the same value as before it was last switched off.
Activates the isoCAL function. The device is automatically adjusted as soon as a
trigger starts the isoCAL function.
Switches the isoCAL function off. This setting change is not available for all
models.
The [isoCAL] button flashes if the balance needs to be adjusted. The isoCAL
function must be manually triggered with the [Adjust] button.
Changes the calibration weight unit to grams.
Changes the calibration weight unit to milligrams.
Explanation
Resets the system settings to the factory default settings.
Deactivates the option of resetting the device menu.
| MCE Models
®