Decommissioning; Decommissioning The Device; Dismantling Device Parts; Dismantling Draft Shield And Components In The Weighing Compartment - Sartorius Cubis MCE Series Operating Instructions Manual

High-capacity micro balance
Hide thumbs Also See for Cubis MCE Series:
Table of Contents

Advertisement

1
11

Decommissioning

11.1

Decommissioning the Device

Procedure
Disconnect the device from the power supply.
t
Disconnect the device from all connections at the installation site.
t
If an accessory is connected to the device: Disconnect the accessory
t
from the device (see instructions for the accessory).
If the upper or lower cover for the rear wall of the weighing compartment
t
is removed: Insert the cover for the weighing compartment into the rear
wall of the weighing compartment (see Chapter 6.5, page 50).
Clean the device.
t
11.2

Dismantling Device Parts

11.2.1
Dismantling Draft Shield and Components in the
Weighing Compartment
Procedure
Remove the weighing pan and the corresponding components from the
t
device (see Chapter 9.1.2, page 66).
Dismount the draft shield (see Chapter 9.1.1, page 65).
t
11.2.2

Dismounting the Connection Cable

Procedure
Place the device on its side (see Chapter 5.5, page 46).
t
Remove the connection cable for the operating display (1) out of the
t
cable channel along the side of the device.
Remove the Ethernet connection cable out of the cable channel along
t
the side of the device.
Place the device back onto its base.
t
Operating Instructions Cubis