Parameters In The "Data Output" Main Menu - Sartorius Cubis MCE Operating Instructions Manual

Semi-micro-, analytical and precision balances
Table of Contents

Advertisement

System Settings
Parameters in the "USB" Submenu
Parameters
Setting values
DEV.USED
NONE*
[Device
Designation]
* Factory setting
7.3.3

Parameters in the "Data Output" Main Menu

Parameters in the "SBI Communication" Submenu
Parameters
Setting values
COM. OUTP.
IND.NO*
IND.AFTER
AUTO W/O
AUTO W/
STOP.AUTO
OFF*
ON
AUTO.CYCL.
EACH VAL*
2. VALUE
INTERV.
FORMAT
16 CHARS
22 CHARS
EXTR.LINE
AUTO.TARE
OFF*
ON
* Factory setting
52
Operating Instructions Cubis
Explanation
Indicates that no device connection is detected at the USB port.
Displays the designation of the devices connected to the USB port.
Explanations
Starts the data output after pressing the key or software command as an
individual value without stability.
Starts the data output after pressing the key or software command as an
individual value after stability.
Activates the automatic data output without stability.
Activates the automatic data output after stability.
Deactivates the option to abort the automatic data output.
The automatic data output is aborted by pressing the [PRINT] key or a
software command.
Starts the automatic data output with a cycle after each value.
Starts the automatic data output with a cycle after every 2nd value.
Starts the automatic data output at the output rate set under "INPUT /
INTERV."
Data output provides 16 characters per line (16 characters for the measured
values only).
Data output provides 22 characters per line (16 characters for the measured
values and 6 characters for identifiers).
Data output provides an additional line with the date, time, and weight
value.
Deactivates automatic taring after data output.
The device automatically tares after data output.
MCE
®