Operation
58
Operating Instructions Cubis
The calibration and adjustment function is executed:
y
— If automatic leveling is configured: The device levels itself
automatically.
— The [CAL.RUN.] display appears and the [Adjust] button flashes.
— The internal adjustment weight is loaded.
— The device is automatically calibrated and the internal calibration
weight is removed.
— The [CAL.END.] display indicates the end of adjustment.
Starting Internal Calibration And Adjustment Manually
Requirements
— For internal calibration: The "INT.CAL." configuration value is set in the
"SETUP " / "BALANCE" / "CAL.ADJ." menu.
— For the adjustment function based on calibration without automatic
adjustment: The "CAL/ADJ." configuration value is set in the
"SETUP " / "BALANCE" / "CAL.SEQ." menu.
— The weighing pan is unloaded.
— The displayed weight value remains unchanged.
Procedure
Press the [Zero] key.
t
If the device cannot be zeroed: Remove the sample to be weighed and
t
re-zero the device.
To start the calibration: Press the [Adjust] button.
t
The calibration and adjustment function is prepared:
y
— If automatic leveling is configured: The device levels itself
automatically.
— The [CAL.RUN.] display appears and the [Adjust] button flashes.
— The internal adjustment weight is loaded.
— The calibration weight value is displayed.
To start the adjustment: Press the [Confirm] button.
t
The device is calibrated and the internal calibration weight is removed.
y
The [CAL.END.] display indicates the end of adjustment.
y
| MCE Models
®