Operating Design
4.8

Parameter List

4.8.1

Parameters in the "SETUP/BALANCE" Menu

Parameter
Settings
AMBIENT
STABLE*
UNSTABL.
V.UNSTBL.
V.STABLE
APP.FILT.
FINAL.RD.*
FILLING
REDUC.
OFF
STABIL.
MAX ACC.
V. ACC.
ACC.*
FAST
V. FAST
MAX.SPEED
* Factory setting
30
Operating Instructions Cubis
Explanation
Sets the ambient conditions to "stable".
Recommended for the following work environment:
— Stable table
— Slight movement in the room
— Slight draft
Sets the ambient conditions to "unstable": Activates the delayed change in
weight values with a reduced output rate.
Recommended for the following work environment:
— Simple office desk
— Room with moving machinery or personnel
— Slight air movement
Sets the ambient conditions to "very unstable": Activates a significantly delayed
change in the weight values and long wait for stability with a further reduction in
the output rate.
Recommended for the following work environment:
— Noticeable and slow floor vibrations
— Noticeable building vibrations
— Weighed goods moved
— Very strong air movements
Sets the ambient conditions to "very stable": Activates a fast change in the
weight values in the event of a load change with a high output rate.
Recommended for the following work environment:
— Very stable table near the wall
— Closed and calm room
— Activates a filter that enables a fast change in the display for very fast load
changes.
— Display changes with minimal load changes (in the digit range) occur more
slowly.
Activates a filter that enables a very fast change in the display with minimal load
changes (e.g. when dosing or filling vessels).
Activates a weak but fast filter that always behaves in the same way for load
changes (e.g. when dosing in automated systems).
Deactivates the active application filter.
Sets the stability to "maximum accuracy".
Sets the stability to "very accurate".
Sets the stability to "accurate".
Sets the stability to "fast".
Sets the stability to "very fast".
Sets the stability to "maximum speed".