Operating Instructions Cubis
8.7.2

Weighing Results with ID Marking

The device and every lot can be assigned an ID number. The ID numbers are
provided in the GLP header during GLP-compliant printing.
Requirements
— In the "DATA.OUTP. / INPUT / ID NO." menu, an ID number is entered for
the device.
— In the "DATA.OUTP. / INPUT / LOT ID" menu, the output of the lot ID line
in the GLP printout is activated.
— In the "DATA.OUTP. / PRNT.PARA. / GLP" menu, the GLP-compliant
printout is activated.
— In the "DATA.OUTP. / PRNT.PARA. / FORMAT" menu, the print output is
set to 22 characters per line.
Procedure
Start the printout. To do so, press the [Print] key.
t
The GLP header is printed with the ID marking set in the menu and the
y
current weight value.
The [GLP] button appears in the operating display.
y
To exit the GLP printout: Press the [GLP] button.
t
The GLP footer is printed.
y
8.8

Starting the Ionization Process

Requirements
— A license key has been entered to use the ionizer.
— The [Ionizer] button appears in the operating display.
Procedure
Press the [Ionizer] button.
t