Digital CMOS Camera C16240-20UP / C16240-20UP01 Instruction manual_Ver.1.3
9-2
PREPARATING FOR IMAGING
Use the following procedure when start operating the camera.
•
When you connect cables, turn off the power supply of the camera and the peripheral devices.
•
When using F-mount (C16240-20UP01), an F-mount adapter can be subject to light leaks due to
the mating flange mechanism. When used with a high sensitivity, long exposure camera, such the
ORCA-Fire, it may be possible to detect photons originating from the F-mount light leakage. In this
case, it is recommended to create a dark condition for the camera such as a dark room, dark box
or wrap the F-mount flange area with a dark cloth. Be careful maintain the clearance distances in
the venting areas of the camera for proper cooling air flow.
1. Connect the equipment as shown in Figure 7-1 or 7-2 before operating of the camera.
2. Turn on the camera.
3. Check cooling fan is operating properly and air is circulating.
9-3
IMAGING
Operate the camera from application software.
9-4
END OF IMAGING
Follow the procedure below when imaging is finished.
1. End the imaging or transmission of image data with the application software.
2. Turn off the camera and peripheral devices.