Digital CMOS Camera C16240-20UP / C16240-20UP01 Instruction manual_Ver.1.3
Trigger Times;
Also in the Synchronous readout trigger mode, synchronous readout can be controlled by
specifying, the number of timing pulses to determine the exposure time. The input trigger is valid
only during the trigger ready is enabled.
The following figure shows the exposure timing when the Trigger Times is set as 4.
External trigger
Exposure time
Sensor readout
CoaXPress data output
Global exposure timing output
Any-row exposure timing output
Trigger ready output (CoaXPress)
40
0(H)
*
1(H)
・
・
・
N-1(H)
N(H)
Figure 11-10 (Trigger Times)
*
* Delay: 4H, Jitter: ≦ 1H
1H readout time : 7.309 μs