The frame rate value is valid 3 digits and rounded down to 4th digit.
The calculation formula and the frame rate value of Start trigger mode (External trigger mode) are
same as Free running mode. About this mode, refer to 10-1-6-3 "Start trigger mode".
2. 8 bit digital output
Operation modes
Free running mode
External trigger mode
(Edge trigger / Level trigger /
Global reset edge trigger /
Global reset level trigger)
External trigger mode
(Synchronous readout trigger)
The frame rate value is valid 3 digits and rounded down to 4th digit.
The calculation formula and the frame rate value of Start trigger mode (External trigger mode) are
same as Free running mode. About this mode, refer to 10-1-6-3 "Start trigger mode".
Digital CMOS Camera C16240-20UP / C16240-20UP01 Instruction manual_Ver.1.3
Binning
Calculation formula
If (Exp2 < 13 682):
1×1
1 / (Vn2 + Vn5) / 1H
Others:
1 / Exp2 / 1H
2×2
4×4
If (Vn2 + Exp2 + 2 < V4):
1×1
1 / (Vn4 + 1) / 1H
Others:
1 / (Vn2 + Exp2 + 3) / 1H
2×2
4×4
1×1
1 / (Vn6 + 1) / 1H
2×2
4×4
Hn × Vn
Max frame rate (fps)
4432 × 2368
31.5
2048 × 2048
79.8
1024 × 1024
264
512 × 512
524
256 × 256
1020
256 × 8
15 200
256 × 4
19 500
2216 × 1184
115
1108 × 592
115
4432 × 2368
31.5
2048 × 2048
79.8
1024 × 1024
262
512 × 512
516
256 × 256
998
256 × 8
10 500
256 × 4
12 400
2216 × 1184
114
1108 × 592
114
4432 × 2368
31.5
2048 × 2048
79.8
1024 × 1024
263
512 × 512
518
256 × 256
1000
256 × 8
11 400
256 × 4
13 600
2216 × 1184
114
1108 × 592
114
31