Digital CMOS Camera C16240-20UP / C16240-20UP01 Instruction manual_Ver.1.3
(4) Global reset Level trigger mode
Global reset function enables to reset the electric charge of all pixels at the same time. Then all
pixels can start exposure at the same time.
The example below is for the trigger level High. With this Global reset Level trigger mode, the
exposure of all pixels begins when the trigger signal becomes High.
External trigger
Exposure time
Sensor readout
CoaXPress data output
USB data output
Global exposure timing output
Any-row exposure timing output
Trigger ready output (CoaXPress)
Trigger ready output (USB)
38
0(H)
*
1(H)
・ 
・ 
・ 
N-1(H)
N(H)
Figure 11-8 (Ex. level High)
*
* Delay: 4H, Jitter: ≦ 1H
1H readout time : 7.309 μs