8. OPERATIONS

8-1

OPERATING PRECAUTIONS

Be careful of the following when you operate the camera.
(1) Cooling method
Cooling of this equipment is done using a Peltier element. With a Peltier element, when
current is supplied, one surface is cooled, and the other surface is heated. The CMOS
image sensor is positioned on the cooling side, and cooling is done by discharging the
heat from the heated surface. This cooling method is passive air-cooling.
(2) Ambient temperature
The recommended ambient temperature for camera operation is between 20 ºC and
25 °C. Thus, the maximum temperatures to which the CMOS image sensor can be
cooled, and the stability of the cooled temperature, are affected by the ambient
temperature. The ambient temperature should be maintained at a constant
temperature in order for cooling to be effective.
(3) Protection circuit
This camera's thermoelectric cooling device is protected by a thermal protection circuit.
If the internal temperature of the camera becomes abnormally hot, the protection circuit
operates to inform the user by a buzzer alarm (beep tone) and lighting the camera's red
LED light while simultaneously cutting the current supply to the Peltier element. As
soon as this protection is implemented, turn off the power switch and unplug the AC
supply. Then remove the cause of the overheating.
8-2

PREPARING FOR IMAGING

Use the following procedure when starting operating of the camera.
(1) Connect devices as shown in Figure 7-1 before you start operation.
(2) Turn on the computer's power switch.
The cooling temperature becomes stable about 5 minutes after cooling begins.
When the cables are connected, confirm the power switch of peripheral device is in the OFF position.
8-3

IMAGING

Start the control and imaging with the application software.
Note
Please refer to the instruction manual attached to the software for the way of using it and the details.
8-4

END OF IMAGING

Carry out the procedure below when imaging is finished.
(1) End the imaging or transmission of image data with the application software.
(2) Turn off the power to the peripheral device.