External trigger
Sensor readout
Read end
Figure 11-23 Programmable timing output (Top to bottom readout)
External trigger
Figure 11-24 Programmable timing output referenced with Hsync (Top to bottom readout)
When you choose Hsync for the reference of programmable timing output, camera can output
some pulses before start the exposure. It is called as Pre-Hsync. You can set the number of Pre-
Hsync.
(3) Trigger ready output
This output behaves the same operation as Normal Area Mode and Lightsheet Readout Mode.
Refer to 10-1-7(4) "Trigger ready output" for the details.
Digital CMOS Camera C16240-20UP / C16240-20UP01 Instruction manual_Ver.1.3
0(H)
1(H)
・ 
・ 
・ 
N-1(H)
N(H)
Vsync
Trigger
Delay
Width
0(H)
1(H)
・ 
・ 
・ 
N-1(H)
N(H)
Hsync
Delay
Width
PreHsync Pulse Num
Delay
Width
Delay
Width