Digital CMOS Camera C16240-20UP / C16240-20UP01 Instruction manual_Ver.1.3
11-2-4 READOUT TIME OF THE HORIZONTAL LINE
Readout time and exposure time can be varied with Lightsheet Readout Mode for synchronizing the
camera readout with the illumination scan.
Available setting range of the exposure time is the following.
Setting range
48
Vn
= Number of vertical lines
1H
= 7.309 µs to 233.9 µs
Readout time = (Vn / 2 + 5) × 1H
Vn
= Number of vertical line
1H
= 7.309 µs to 233.9 µs
1H to Vn / 2 × 1H