11-1-4 FRAME RATE CALCULATION
(1) CoaXPress
Exp1 = 7.309 µs to 10 s (input in units of seconds)
1H = 7.309 µs
Exp2 = ROUNDUP (Exp1 / 1H) * Round up to integer
Vn1 = Number of vertical lines
Vn2 = (Vn1 / 2) + 5
Conditions
Exp2 < 13 682
Exp2 ≥ 13 682
Vn4 = MAX (Vn2 + 2, Exp2)
Operation modes
Free running mode
External trigger mode
(Edge trigger / Level trigger /
Global reset edge trigger /
Global reset level trigger)
External trigger mode
(Synchronous readout trigger)
The frame rate value is valid 3 digits and rounded down to 4th digit.
The calculation formula and the frame rate value of Start trigger mode (External trigger mode) are
same as Free running mode. About this mode, refer to 10-1-6-3 "Start trigger mode".
The calculation formula and the frame rate value do not depend on the bit depth of digital output.
Digital CMOS Camera C16240-20UP / C16240-20UP01 Instruction manual_Ver.1.3
Vn3
Exp2 < Vn2
0
Exp2 ≥ Vn2
Exp2 - Vn2 + 1
0
Calculation formula
If (Exp2 < 13 682):
1 / (Vn2 + Vn3) / 1H
Others: 1 / Exp2 / 1H
1 / (Vn2 + Exp2 +3) / 1H
1 / (Vn4 + 1) / 1H
Horizontal
Vertical
Max frame rate (fps)
2368
2048
1024
4432
512
256
8
4
2368
2048
1024
4432
512
256
8
4
2368
2048
1024
4432
512
256
8
4
115
132
264
524
1020
15 200
19 500
114
132
262
516
998
10 500
12 400
114
132
263
518
1000
11 400
13 600
29

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orca-fire c16240-20up01

Table of Contents