Table 7. Settings for Fiber Tests(cont.)
Test Limit
Select the correct test limit for the job. To see a different
group of limits, tap MORE, then tap the name of a group.
To make a custom limit, tap Custom in the Limit Groups list.
See the Technical Reference Handbook.
Reference
On the No. of Connectors/Splices screen, set the number of
Method
jumpers you will use in each fiber path when you set the
reference. The dotted lines in the diagram on the screen
show you which parts of the link are included in the test
results.
The number of jumpers you use has these effects on loss
measurements:
1 Jumper: Loss measurements include the connections at
both ends of the link. The figures in this manual show 1
Jumper connections.
2 Jumper: Loss measurements include one connection at
one end of the link.
3 Jumper: Loss measurements do not include the
connections at the ends of the link. The tester measures
only the loss of the fiber.
This setting does not change the loss measurements, but it
can change the PASS/FAIL result for test limits that use a
calculated loss limit. For all test limits, the tester saves this
setting to show the reference method you used.
Most cable manufacturers will give you a warranty
on a fiber installation only if you use the 1 Jumper
reference method when you certify the
installation.
Different standards use different names for the
three methods. See Appendix A.