Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
Table of Contents Title Page Introduction ......................1 Contact Fluke ..................... 1 Safety Information ....................2 Symbols ......................3 Specifications ..................... 4 AC Voltage ..................... 4 General ......................4 Required Equipment ..................5 Performance Tests ..................... 5 Test the Display ..................... 5 Backlight ......................
Page 7
List of Figures Figure Title Page All Segments of the Display ................... 5 Performance Test Connections ................7 Calibration Password Reset ................... 9 Battery Replacement ....................12...
Read "Safety Information" before you use the Product. This manual has the verification and calibration adjustment procedures for the v3001 FC Wireless DC Voltage module (the Product). Please see the v3001 FC Quick Reference Guide for usage information. Contact Fluke To contact Fluke, call one of the following telephone numbers: •...
FC Calibration Manual Safety Information A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test. Warning To prevent possible electrical shock, fire, or personal injury: Carefully read all instructions.
Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation” product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
FC Calibration Manual Specifications Accuracy is specified as ±( [% of reading] + [number of least significant digits] ). All ranges are autoranging. Accuracy is specified from 1 % to 100 % of the range obtained by autoranging, from 18 °C to 28 °C.
2. Push while you keep pushed until all of the display segments are shown. See Figure 1. hby002.eps Figure 1. All Segments of the Display If segments of the display are missing, repair is necessary. See “Contact Fluke”.
To verify that the keypad functions, turn on the Product and push each button separately. Each button push will turn on a display annunciator. If the buttons do nothing, repair is necessary. See “Contact Fluke”. DC Voltage Test Before you do the dc voltage test: 1.
Wireless DC Voltage Before Calibration Adjustment hcc07.eps Figure 2. Performance Test Connections Before Calibration Adjustment Before the Product calibration can be adjusted, you must go through the maintenance mode menu and enter your password. Maintenance Mode The Product maintenance mode can be used to set different parameters on the Product that include auto power off, backlight adjustment, and calibration.
FC Calibration Manual Password Entry To go to the calibration mode, push five times until CAL is shown. You will need to enter a password to access calibration mode. To enter the password: 1. Push and the CAL counter is shown. For example n002.
Wireless DC Voltage Before Calibration Adjustment Restore the Default Password If the calibration password is lost, the default password (1234) can be manually restored with the subsequent steps: Warning To prevent electric shock or personal injury, remove all input signals before you open the Product.
FC Calibration Manual Calibration Adjustment The Product features closed-case calibration adjustment and uses known reference sources. The Product measures the applied reference source, calculates correction factors, and stores the correction factors in nonvolatile memory. Should the Product fail any of the performance tests, do the calibration adjustment procedure.
Wireless DC Voltage Maintenance Maintenance Clean the Product Caution To prevent possible damage to the Product or to equipment under test, do not use abrasive cleaners. They will damage the case. To clean the Product, use a cloth with a mild cleaning solution. Battery Replacement ...
LARGE ALLIGATOR CLIP BLACK, BULK variable LARGE ALLIGATOR CLIP RED, BULK 376756 Battery, AA 1.5 V, NEDA 15 A, IEC LR6 4466335 INFORMATION PACK, FLK-V3000/V3001 FC [1] See www.fluke.com for more information about the test leads and alligator clips available for your region.