Versiv Series Cabling Certification Product Family
Users Manual
76
Find Quality Products Online at:
Set the reference only after the testers are at an
ambient temperature between 10 °C and 40 °C
(50 °F and 104 °F).
The tester will not let you set the reference if the
patch cord is longer than 30 cm (12 in).
You can also set the reference with a 50 Ω patch
cord.
On the home screen, select a coaxial cable test.
3
On the home screen, tap TOOLS, then tap Set Reference.
4
On the SET REFERENCE screen tap TEST.
5
Coaxial adapter
with F connector
adapter
DSX-CHA003
COAX ADAPTER
Figure 28. Reference Connections for Tests on Coaxial Cabling
GlobalTestSupply
www.
Notes
30 cm (12 in) 75 Ω
coaxial patch cord
Coaxial adapter
with F connector
adapter
DSX-CHA003
COAX ADAPTER
.com
GPU179.EPS