4 Installation

4.1 Site requirements

4.1
Site requirements
Space requirements
Prepare a clean working area on a stable laboratory bench. The bench must comply
with the specifications in the following table.
Parameter
Minimum bench area for operating
Amersham Imager 680
Free space required around the imager
Load capacity
Inclination of bench surface
74
WARNING
Access to power cord. Do not block access to the power cord. The
power cord must always be easy to disconnect.
NOTICE
This equipment is not intended for use in residential environments
and may not provide adequate protection to radio reception in
such environments.
Specification
70 x 60 cm
At least 40 cm free space in front of the
instrument
10 cm free space on all other sides
980 N (100 kg) or higher
Horizontal ±2°
Amersham Imager 680 Operating Instructions 29271308 AB