1.8 CLEANING AND MAINTENANCE

After each use, clean the defibrillator and the reusable pads using a soft, damp cloth moistened with any of
the following solvents:
Water and soap.
Chlorhexidine and water mixture (30 ml Chlorhexidine /liter of water).
Ammonia.
Hydrogen peroxide.
If necessary, sterilize just the defibrillation pad part touching the patient skin only with liquid CIDEX.
D
O NOT IMMERSE ANY PART OF THE DEFIBRILLATOR IN FLUIDS
D
O NOT LET ANY FLUID ENTER THE CASE OF THE DEVICE
LCD
.
DISPLAY
D
O NOT STERILIZE THE DEVICE
D
O NOT REUSE THE DISPOSABLE PADS
P
OSSIBLE EQUIPMENT DAMAGE
D
O NOT CLEAN ANY PART OF THIS DEVICE OR ACCESSORIES WITH BLEACH
ABRASIVE OR FLAMMABLE CLEANING AGENTS
SPECIFIED IN ACCESSORY
The operator has to do daily maintenance checks that will help ensure that the device stays in perfect
operational condition.
Check the case of the device for any apparent damage.
Check the ports (defibrillator lead port, patient cable port, AC plug and cable, paddles).
Check the accessories, especially the defibrillation pads and cables, to see that they are in good condition.
Check the battery status and if the level is low attach the power cord to the AC line. The battery is
rechargeable and intended to be used for standby operation. The defibrillator automatically switches to
battery power when the power cord is disconnected from an AC outlet or from the defibrillator.
Rescue Life
1 - INTRODUCTION
D
.
.
.
. D
O NOT ATTEMPT TO STERILIZE THIS DEVICE OR ANY ACCESSORIES UNLESS OTHERWISE
O
I
.
PERATING
NSTRUCTIONS
9
– Instruction For Use
.
O NOT USE ABRASIVE MATERIALS IN CLEANING THE UNIT
,
BLEACH DILUTION
1-31
Rev.0.5 – 30/03/2022
,
ESPECIALLY ON THE
,
. D
OR PHENOLIC COMPOUNDS
O NOT USE