Technical Features - Progetti Rescue Life 9 Instructions For Use Manual

Defibrillator
Table of Contents

Advertisement

APPENDIX C

TECHNICAL FEATURES

Software version: V3.0
ECG Monitoring
• CMRR > 98 dB
• Input impedance: 50 Mohm
• Patient connection:
Defibrillation pads and 3, 5 or 10 wire ECG cable.
• Bandwidth:
0.6 to 40 Hz (-3 dB) in monitor mode.
0.05 to 120 Hz (-3 dB) in diagnostic mode.
• ECG trace parameters:
Speed: 5, 10, 25, 50 mm/sec.
Gain: 2.5, 5, 10, 20, 40 mm/mV with patient cable. AUTO with pads.
Alarm: HR max settable = 250 bpm, HR min settable = 20 bpm.
Filters: 50/60 Hz, EMG filter, base line.
Trace: I or II or III with 3 wires patient cable
Traces: 3+3+1 (I, II, III – aVR, aVF, aVL, V) with 5 wires patient cable.
Traces: 3+3+6 (I, II, III – aVR, aVF, aVL – V1 to V6) with 10 wires patient cable.
• Heart rate:
Digital readout on the display from 20 a 300 bpm (± 5% or ±3 bpm, on higher value).
Defibrillator
• Operation mode:
Manual, Advisory, automatic AED.
• Defibrillable impedance:
Compensated from 25 ohm to 200 ohm.
• Manual Mode:
Syncro/Asyncro.
• Defibrillation pads:
Standard or disposable, adult and pediatric.
8-1