RESCUE LIFE 9 CHECKLIST

PROGETTI S.r.l. recommend using the following checklist to monitor the status of RESCUE LIFE 9. It includes
all the tests for checking the functionality and the safety.
Operator should test the defibrillator at least once a week.
It is possible to do a discharge test by connecting the test load (optional) using the cable adapter for
disposable pads.
CONDITION
Check integrity of the enclosure and the keys and make sure the unit is clean.
ACCESSORIES
Check and clean the accessories. Check the integrity of the power, ECG, SPO2, TEMP,
ETCO2, NIBP cable and sensor. Check expiration date of DISPOSABLE MULTIFUNCTION
ELECTRODES.
AUTO TEST
Perform auto test as explained in the section (page 65)
ELECTRODES
Check the integrity of the ELECTRODES, the connector and the cable. Make sure the
ELECTRODES are clean and free of traces of gel. To test the paddles switch on
defibrillator, set an energy of 1 Joule, charge the capacitor by pressing the buttons on
the paddles and release the discharge into the air, taking care not to touch the paddles
together.
VOICE
Switch on defibrillator, select ADVISORY mode from the menu and verify that voices
are audible.
AED
If you have the test load (optional) it is possible to simulate a condition of asystole,
and verify that the AED module works properly, recognizing that condition and
recording data.
Turn on the defibrillator, connect the load test and select automatic (AED) by the
menu. Wait for the defibrillator recognizes asystole and the recording coming to an
end.
Switch off defibrillator then check in Data Base if record was recorded properly.
PRINT
Switch on the defibrillator and press PRINT key. Print will start. Press PRINT key again
to stop print. Check the quality of the printed report.
5-1
OK KO
OK KO
OK KO
OK KO
OK KO
OK KO
OK KO