IMPENDANCE LIMITS

RESCUE LIFE 9 does not release the shock if the patient impedance is less than 25 ohm or over 200 ohm.

SYNC/NO SYNC MODE

When the RESCUE LIFE 9 is switched on it is automatically set on no-sync mode.
Only the operator can set the sync/no-sync mode which is clearly displayed on the screen. The device can
not automatically set the sync mode.
In the sync mode the device releases the defibrillation shock only when the 'R' peak in the 'QRS' complex is
detected.
The maximum response time between the "R" peak and the defibrillation shock is less than 60ms.

CHARGING TIME TO ACHIEVE THE MAXIMUM ENERGY (230 JOULE)

When the device is connected to the AC supply (nominal AC voltage) and batteries are 100% charged :< 8
sec.
When the device is connected to the AC supply (AC voltage 90%) after 15 shocks :< 10 sec.
7-2