Etco 2 Module Specifications - Progetti Rescue Life 9 Instructions For Use Manual

Defibrillator
Table of Contents

Advertisement

3 – ACQUISITION OF BIOMEDICAL PARAMETERS
After using the CO
sensor, clean the CO
2
liquids and leave it to dry. Clean and disinfect the other parts of the CO
liquids:
Ethanol (15°C (59 F), 76.9 to 81.4% by vol)
Diluted mild detergent
You cannot clean, disinfect or sterilize them. Immediately replace them with ones when they become dirty.
3.3.7 ETCO
MODULE SPECIFICATIONS
2
Measuring method
Measuring range
Measuring accuracy
Accuracy stability
Total system response time
Powe source
Detectable respiration rate
Rise time
EtCO
calculation
2
Data sampling rate
Warm-up time
Degree of protection provided by
enclosures
Degree of protection against shock
Degree of protection against
electrical shock
Degree of safety of application in
the presence of FLAMMABLE
ANAESHETIC MIXTURE WITH AIR, OR
OXYGEN, OR NITRUS OXIDE
Mode of operation
Applicable patients (weight)
Operating environment
Rescue Life
sensor with a cotton swab moistened with any of the following
2
NOTE: When using flammable solvent such as ethanol for cleaning and
disinfecting, ventilate the room adequately.
Mainstream
0 to 20 kPa (0 to 150 mmHg)
± 0.27 kPa (0≤ CO
± 2 mmHg (0≤ CO
± 5% of gas level (5.33 < CO
± 5% of gas level (40 < CO
± 7% of gas level (9.33 < CO
± 7% of gas level (70 < CO
± 10% of gas level (13.3 < CO
± 10% of gas level (100 < CO
(noncondensing)
Measurement accuracy is guaranteed for 6 hours after power on
≤ 0.5 s
3.3 V
± 5% to 5 V
DC
0-150 breaths/min ± 1 breath/min
< 60 ms for 10% to 90%
Calculated from the maximum CO
40 Hz
about 10 s
Sensor part: IP67 (not protected during use)
Adapter part: IP34
MIL STD 810G: 2008 516.6 4.6.5 (except interface connector)
Shock resistance: 100 drops from 2 m (except interface connector)
Defibrillator-proof type BF applied part
Equipment not suitable for use in the presence of FLAMMABLE
ANAESHETIC MIXTURE WITH AIR, OR OXYGEN, OR NITRUS OXIDE
CONTINUOUS OPERATION
≥ 7kg (Adult) with YG-211TW airway adapter
< 7kg (Infant/neonatal) with YG-213TW or YG-214TW airway
adapter. In this case the endotracheal tube size must be 4.0 mm or
less
Temperature: 0 to 40°C (32 to 104 F)
Humidity: 15 to 95% RH (noncondensing)
Atmospheric pressure: 70 to 106 kPa
9
– Instruction For Use
sensor with any of the following
2
≤ 5.33 kPa)
2
≤ 40 mmHg)
2
≤ 9.33 kPa)
2
≤ 70 mmHg)
2
≤ 13.3 kPa)
2
≤ 100 mmHg)
2
≤ 20 kPa)
2
≤ 150 mmHg)
2
± 5%
DC
in expiration
2
3-27