3 – ACQUISITION OF BIOMEDICAL PARAMETERS
4. Press the KNOB to confirm the parameter to change. It will be shown in yellow;
5. Rotate the KNOB to change the value (default superior limit: 250 bpm, default inferior limit: 30 bpm);
6. Press the KNOB to confirm;
7. Rotate the KNOB to change another parameter;
8. Select "EXIT" to save and exit.
3.1.4 ECG PERFORMANCE (REF. §201.7.9.2.9.101 OF EN 60601-2-27:2016)
Heart Rate averaging (ref. b3):
- Heart Rate is averaged over the last 3 'R' peaks detected.
- Heart Rate update depends on the actual 'R' peaks repetition period (example: for 60 bpm the update is 3
seconds).
Heart Rate accuracy (ref. b4): (Fig. 201.101 of EN 60601-2-27)
- Waveform A1 (Ventricular bigeminy): HR = 40 bpm, larger R peaks counted
- Waveform A2 (Slow alternating ventricular bigeminy): HR = 30 bpm, only large complexes counted