2.1.6.2 "ADV" MODE
DESCRIPTION
The "ADV" (Advisory) mode allows the user to:
•
do the same operations as the "Manual" mode;
•
establish if a defibrillation discharge is needed or not following messages shown on the
"MESSAGES" box of the display ("SHOCK ADVISED" or "NO SHOCK ADVISED") or played as sound.
DEFIBRILLATION PROCEDURE IN ADV MODE
1. POWER ON
See "Manual" mode.
2. DEFIBRILLATION ELECTRODES CONNECTION
See "Manual" mode.
3. SET SYNCHRONISM OR NO SYNCHRONISM
See "Manual" mode.
4. SET ECG SOURCE
See "Manual" mode.
5. ATTACHMENT OF ECG SENSOR TO THE PATIENT
See "Manual" mode.
6. "ADV" MODE SELECTION
Press "DEF MODE" function to change from "MANUAL" (default) to "ADV".
9
Rescue Life
– Instruction For Use
2 - THERAPY
2-12
Rev.0.5 – 30/03/2022