1.10 BATTERY CHARGING

When the battery charge level is 10%, the message "BATTERY LOW. AUTO SHUT-OFF IN: 120sec. CHARGE /
SHOCK DISABLED" is shown on the display and a pulsed sound is played, so the battery or batteries need to
be recharged. The value of countdown seconds (sec) is updated every 10 seconds. After 120 seconds the
defibrillator will do shut-off automatically.
After this warning it is no longer possible to perform defibrillation shocks..
Connect the defibrillator to the AC line to start the battery or batteries recharge and to continue the use of
defibrillator.
Tipically, new fully depleted batteries recharge time is 2.5 hours (per battery) to regain full capacity.
When the charging is finished BATTERY LED will turn off.
BATTERY LED status
statement A
statement B
W
HEN THE DEVICE IS IN POWER
I
F AFTER THIS TIME THE CHARGING LIGHT DOES NOT GO OFF
CHANGING THE BATTERIES
9
Rescue Life
1 - INTRODUCTION
-
,
OFF
DO NOT LEAVE THE
.
1-35
– Instruction For Use
DESCRPTION
When the orange BATTERY LED is ON it informs
that the battery are in charging (by connection
to AC line).
When the battery charge is finished the
BATTERY LED is OFF. Only AC POWER LED will be
ON until AC line will be disconnected.
AC POWER CABLE
CONNECTED MORE THAN
,
PLEASE CONTACT THE SERVICE CENTER FOR