HandyLab 200
Selecting the
displayed
measured parameter
Stability control
(AutoRead )
ba77063e03
01/2018
4.4.2 Measuring
You can carry out conductivity measurements as follows:
1
Perform the preparatory activities according to section 4.4.1.
2
Immerse the conductivity measuring cell in the test sample.
ϰ
1417
25.0
[Tr25][nLF][ 0.475 1/cm]
15.03.2014
Menu
10:00
You can switch between the following displays with <M>:
Conductivity [μS/cm] / [mS/cm]
Resistivity [Ω·cm] / [kΩ·cm] / [MΩ·cm]
Salinity SaL [ ]
Total dissolved solids TDS [mg/l] / [g/l]
The factor to calculate the total dissolved solids is set to 1.00 in the fac-
tory. You can adjust this factor to meet your requirements in the range
0.40 ... 1.00. The factor is set in the Measurement menu for the param-
eter, TDS.
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of
the setting for automatic Stability control (see page 23) in the Measure-
ment menu.
1
Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
µS/cm
°C
Operation
25