Operation
Settings
28
The settings are made in the Measurement menu for the measured
parameter, conductivity. To open the settings, display the required
parameter in the measured value display and press the <F1>/[menu] or
<ENTER> key. After completing the settings, switch to the measured
value display with <M>.
Menu item
Calibration /
Calibration record
Calibration / Calibra-
tion data storage
Calibration /
Calibration interval
Measurement /
Measuring cell /
Type
Possible
Explanation
setting
-
Displays the calibration record
of the last calibration.
-
Displays the last calibration
records.
1 ... 999 d
Calibration interval for the
measuring cell (in days).
The meter reminds you to cali-
brate regularly by the flashing
sensor symbol in the measur-
ing screen.
Measuring cell used
Cal
Measuring cells the cell con-
stant of which is determined by
calibration in the KCL control
standard solution.
Calibration ranges:
0.450 to 0.500 cm
to 0.880 cm
The currently valid cell con-
stant is displayed in the status
line.
LF313T
Measuring cell LF313T,
nominal cell constant
0.100 cm
The cell constant can be
adjusted in the range from
0.090 to 0.110 cm
LF213T
Measuring cell LF213T,
nominal cell constant
0.010 cm
The cell constant is perma-
nently set.
HandyLab 200
-1
and 0.800
-1
-1
.
-1
.
-1
.
ba77063e03
01/2018