HandyLab 200
Application tips
Overview
ba77063e03
01/2018
Nonlinear temperature compensation (nLF) according to EN 27
888
Linear temperature compensation (Lin) with adjustable coeffi-
cients in the range of 0.000 ... 10.000 %
No temperature compensation (off)
Note
The reference temperature and temperature compensation are set in
the Measurement menu for the parameter, conductivity (see section
4.4.4).
Select the following temperature compensations given in the table
according to the respective test sample:
Test sample
Natural water
(ground water, surface
water, drinking water)
Ultrapure water
Other aqueous solu-
tions
Salinity (seawater)
4.4.4 Settings for conductivity measuring cells
The following settings are possible for conductivity measuring cells:
Calibration record (display)
Calibration interval
Measuring cell/cell constant
Reference temperature
Temperature compensation
TDS factor
Unit of the temperature
Automatic Stability control
/
K
Temperature compensation
nLF
according to EN 27 888
nLF
according to EN 27 888
Lin Set linear
temperature coefficient
0.000 ... 10.000 %/K
Automatic nLF according to
IOT (International Oceano-
graphic Tables)
Operation
Display
indica-
tor
nLF
nLF
Lin
Sal, nLF
27