Xylem SI Analytics HandyLab 200 Operating Manual page 41

Conductivity meter
Table of Contents

Advertisement

HandyLab 200
Cell constants
Reference temperature
Accuracy (± 1 digit)
ba77063e03
01/2018
ϰ [mS/cm]
ρ (Resistivity)
[Ohm*cm]
ρ (Resistivity)
[kOhm*cm]
ρ (Resistivity)
[MOhm*cm]
SAL
TDS
T [°C]
T [°F]
* only possible with cells of the cell constant, 0.010 cm
** only possible with cells of the cell constant, 0.010 cm
Cell constant C
To be calibrated
Adjustable
Reference tempera-
ture
Adjustable
Parameter
ρ
ϰ and
/ temperature compensation
None (Off)
2.00 ... 19.99
20.0 ... 199.9
200 ... 1000
1.000 ... 1.999
2.00 ... 19.99
20.0 ... 199.9
200 ... 1999
2.00 ... 19.99
20.0 ... 199.9
200 ... 1999
2.00 ... 19.99**
20.0 ... 199.9*
0.0 ... 70.0
according to the IOT table
0 ... 1999 mg/l
2.00 ... 19.99 g/l
20.0 ... 199.9 g/l
-5.0 ... +105.0
+23.0 ... +221.0
Values
-1
0.450 ... 0.500 cm
-1
0.800 ... 0.880 cm
-1
0.010 cm
(fixed)
-1
0.090 ... 0.110 cm
-1
0.250 ... 25.000 cm
Values
20 °C (Tref20)
25 °C (Tref25)
Accuracy
Temperature of the test
sample
± 0.5 %
Technical data
0.01
0.1
1
0.001
0.01
0.1
1
0.01
0.1
1
0.01
0.1
0.1
1
0.01
0.1
0.1
0.1
-1
-1
-1
or 0.090 ... 0.110 cm