Command Description
Secondary Menu
Read proportional band setting
Set proportional band to n
Read cutout setting
Set cutout setting:
Set cutout to n degrees
Reset cutout now
Read heater power
(duty cycle)
Configuration Menu
Probe Menu
Read R0 calibration parameter
Set R0 calibration parameter to
Read ALPHA calibration
parameter
Set ALPHA calibration parameter
n
to
Serial Interface Menu
Read serial sample setting
Set serial sampling setting to n
seconds
Set serial duplex mode:
Set serial duplex mode to full
Set serial duplex mode to half
Set serial linefeed mode:
Set serial linefeed mode to on
Set serial linefeed mode to off
Table 5. Interface Command Summary (cont.)
Command Format
pr[op-band]
pr[op-band]=n
c[utout]
c[utout]=n/r[eset]
c[utout]=n
c[utout]=r[eset]
po[wer]
r[0]
n
r[0]=n
al[pha]
al[pha]=n
sa[mple]
sa[mple]=n
du[plex]=f[ull]/h[alf]
du[plex]=f[ull]
du[plex]=h[alf]
lf[eed]=on/of[f]
lf[eed]=on
lf[eed]=of[f]
Command
Returned
Example
pr
pb: 9.999
pr=0.326
c
cu: 999 {x},{xxx}
c=120
c=r
po
po: 9999
r
r0: 999.999
r=100.324
al
al: 9.9999999
al=0.0038433
sa
sa: 9999
sa=0
du=f
du=h
lf=on
lf=of
Calibration Bath
Digital Communication Interface
Returned
Example
pb: 0.315
cu: 120 C, in
po: 1
r0: 100.578
al: 0.0038573
sa: 1
Acceptable
Values
Depends on
Configuration
Temperature
Range
98.0 to 104.999
.00370 to
.0039999
0 to 4000
FULL or HALF
ON or OFF
45