7381
User's Guide
Problem
The displayed process
temperature is in error and the
controller remains in the cooling or
the heating state at any set-point
value
54
Table 6. Troubleshooting (cont.)
Possible causes may be either a faulty control probe or
erroneous data in memory.
The probe might be disconnected, burned out, or
shorted. The Probe is insided the stirrer motor cover.
Make sure that the probe is connected properly.
Check the probe with an ohmmeter to see if it is
open or shorted. The probe is a platinum 4-wire Din
43760 type. The resistance should read 0.2 Ω to
2.0 Ω between pins 1 and 2 on the probe connector
and 0.2 Ω to 2.0 Ω between pins 3 and 4. The
resistance should read from 100 Ω to 300 Ω
between pins 1 and 4 depending on the temperature.
If the probe is defective, contact Fluke Calibration.
If the problem is not the probe, erroneous data in
memory can be the cause. Re-initialize the memory
as discussed in the problem The display flashes
out
and an incorrect process temperature. If the
problem remains, the cause may be a defective
electronic component, contact Fluke Calibration.