Calibration Procedure; Calibration Points - Fluke 7381 User Manual

Calibration bath
Table of Contents

Advertisement

Calibration Procedure

Consider this procedure a general guideline. Each laboratory should write their own
procedure based on their equipment and their quality program. Each procedure should
have an uncertainty analysis based on the laboratory equipment and environment.
In some instances, you may want to calibrate the Bath to improve the temperature set-point
accuracy. To calibrate, adjust the controller probe calibration constants R
temperature of the Bath (measured with a standard thermometer) agrees more closely with the
Bath set-point. The thermometer used must be able to measure the bath fluid temperature with
higher accuracy than the desired accuracy of the Bath. Use a good thermometer and carefully
follow procedures to calibrate the Bath to an accuracy of better than 0.02 °C over a range of
100 degrees.

Calibration Points

When you calibrate the Bath, adjust R
different bath temperatures. Use any two reasonably-separated bath temperatures for the
calibration. However, best results are obtained when you use bath temperatures that are just
within the most useful operating range of the Bath.
The farther apart the calibration temperatures, the greater the calibrated temperature
range, and the calibration error.
If, for example, 0 °C and 100 °C are chosen as the calibration temperatures, the Bath can
achieve an accuracy of ±0.3 °C over the range –10 °C to 110 °C. Choosing 30 °C and 70 °C
allows the Bath to possibly have a better accuracy of ±0.1 °C over the range 25 °C to 75 °C, but
outside of that range, the accuracy may be only ±0.5°C.
Note
and ALPHA to minimize the set-point error at each of two
0
Note
Calibration Bath
Calibration Procedure
and ALPHA so that the
0
47

Advertisement

Table of Contents
loading

Table of Contents