 Low battery indicator AP: When the unit is turned on, the
«battery» icon will keep flashing to remind the user to replace
the battery.
6. Setting Date, Time and Beeper Functions
Setting the date and time
1. After the new battery is fitted, the year number flashes in the
display BQ. You can set the year by pressing the M-button 3.
To confirm and then set the month, press the MODE button 4.
2. Press the M-button 3 to set the month. Press the MODE
button 4 to confirm and then set the day.
3. Follow the previously mentioned instructions to set the day,
hours and minutes.
4. Once you have set the minutes and pressed the START/IO
button 6, the date and time are set and the time is displayed.
If no button is pressed for 20 seconds, the device automat-
ically switches to ready for measuring 9.
Cancel time setup: Press the START/IO button 6 during
time setup. The LCD will show Date/Time icons with «--:--».
After that press the START/IO button 6 to start the measure-
ment. If no further action is taken within 60 seconds, the
device will automatically turn off.
Change current date and time: Press and hold the MODE
button 4 for approx. 8 seconds until the year number starts
to flash BQ. Now you can enter the new values as described
above.
Setting the beeper
1. Press and hold the MODE button 4 for 3 seconds to set the
beeper BR.
2. Press the M-button 3 to either turn the beeper on or off. The beeper
is activated when the beeper icon BR is shown without a cross.
When the beeper setting has been chosen, press the
START/IO button 6 to enter the «ready for measuring»
mode; otherwise the device automatically switches to ready
for measuring after 10 seconds 9.
7. Changing between Body and Object Mode
1. Press the START/IO button 6. The display 2 is activated to
show all segments for 1 second.
2. The default mode is body mode AL. Press the MODE button 4
to switch to object mode AM. For switching back to body mode,
press the MODE button again.
IR 210
8. Directions for Use

Measuring in body mode

Important: Before each measurement, fit a new undamaged probe
cover CK on the measuring sensor 1. Failure to do so will result in
incorrect temperature measurements. How to correctly fit a new
probe cover CL is pictured at the beginning of this instruction.
1. Press the START/IO button 6. The display 2 is activated to
show all segments for 1 second.
2. When the «°C» or «°F» icon is flashing, a beep sound is heard
and the thermometer is ready for measuring 9.
3. The probe LED light is activated and will keep flashing.
4. Straighten the ear canal by pulling the ear up and back to give
a clear view of the eardrum.
 For children under 1 year; pull the ear straight back.
 For children of 1 year to adult; pull the ear up and back.
Also refer to the short instructions at the front.
5. While gently pulling the ear, insert the probe snugly into the ear
canal.
6. The probe LED light will stop flashing (stays illuminated) and
«good» will be displayed on the LCD, when the measuring
sensor detects an appropriate position.
7. Immediately press the START/IO button 6. Release the button
and wait for the beep sound. This indicates the end of the
measurement.
8. Remove the thermometer from the ear canal. The display shows
the measured temperature AK.
9. Replace the probe cover CK before starting a new measurement.
10.For the next measurement wait until the «°C»/«°F» icon is
flashing and follow steps 5-7 above.
11.Press and hold the START/IO button 6 for 3 seconds to turn
off the device; otherwise the device will automatically switch off
after approx. 60 seconds.

Measuring in object mode

Important: Remove the probe cover before each measurement in
object mode. Failure to do so will result in incorrect temperature
measurements.
1. Press the START/IO button 6. The display 2 is activated to
show all segments for 1 second.
2. Press the MODE button 4 to switch to object mode.
3. Aim the thermometer at the center of the object you want to
measure with a distance of no more than 5 cm. Press the