This Digital Antimicrobial Medical Thermometer provides highly accurate readings over the human body temperature range. At the same time, it reduces the microbial flora and minimizes the dispersion of contagious microorganisms, providing high safety to the user.
Surfaces made or covered by special copper alloys, have strong antimicrobial properties against a wide variety of microorganisms*. Copper alloys emit antimicrobial copper ions Cu+ that whilst in contact with microbes and bacteria rupture their cellular membranes, thus destroying these microorganisms. This activity reduces the microbial flora on the coated area and – due to the «halo phenomenon» – simultaneously causes a drastic reduction in pathogens on the remaining body of the thermometer. Thermometers containing parts with antimicrobial copper alloys drastically reduce microbial flora, minimizing the dispersion of contagious microorganisms and thus providing high safety to the end user.
* Efstathiou A. Panos «The Role of Antimicrobial Copper Surfaces in Reducing Healthcare associated Infections», European Infectious Disease, Volume 5, Issue 2, Autumn 2011.
To turn on the thermometer, press the ON/OFF button ; a short beep signals «thermometer ON». A display test is performed. All segments should be displayed.
The last measurement reading will be shown on the display automatically for 2 seconds with the «M» icon.
Then at an ambient temperature of less than 32°C, an «L» and a flashing «°C» appear at the display field . The thermometer is now ready for use.
Correct functioning of the thermometer is tested automatically each time it is turned on. If a malfunction is detected (measurement inaccuracy), this is indicated by «ERR» on the display, and a measurement becomes impossible. In this case, the thermometer must be replaced.
Before use, keep this thermometer out of any physical contact, for at least 2 hours. This utilizes the antimicrobial copper properties, as described in chapter «Antimicrobial Copper Properties». The use of antimicrobial copper is a supplement to and not a substitute for standard infectious control practices; Users must continue to follow all current infection control and cleaning practices. We recommend cleaning the thermometer as described in chapter «Cleaning and Disinfecting». Choose the preferred measuring method. When taking a measurement, the current temperature is continuously displayed and the «°C» symbol flashes. If the beep is heard 10 times and the «°C» is no longer flashing, the predictive end-temperature has been determined and the thermometer can be read now.
10 short beeps willsound when the temperature is higher than 37.5°C in order to alert the patient that he/she may have fever.
Reference: Oral temperature.
To achieve comparable results allow a 1 minute interval time between measurements.
To prolong the battery life, turn off the thermometer by briefly pressing the ON/OFF button . Otherwise the thermometer will automatically turn off after about 10 minutes.
If the ON/OFF button is pressed for more than 3 seconds when turning on the thermometer, the automatically stored maximum temperature during the last measurement will be displayed. At the same time, a «M» for memory will appear on the display. About 3 seconds after the button is released, the temperature value disappears and the thermometer is ready for measurement.
Clean the thermometer with a soft, dry cloth or with a cotton tissue moistened with Isopropyl alcohol (70%). Don't let the thermometer come into contact with any chemical thinner!
Please never immerse into liquids!
When the «» symbol (upside-down triangle) appears at the display field, the battery is flat and needs replacing. To replace the battery remove the battery compartment cover
from the thermometer. Insert the new battery with the + at the top. Make sure you have a battery of the same type to hand. Batteries can be purchased at any electrical store.
Type: | Predictive maximum thermometer |
Measurement range: | 32.0°C to 42.9°C Temp. < 32.0 °C: display «L» for low (too low) Temp. > 42.9 °C: display «H» for high (too high) |
Measurement accuracy: | ± 0.1°C between 34°C and 42°C |
Operating conditions: | 10 - 40°C; 15-95% relative maximum humidity |
Storage conditions: | -25 - +60°C; 15-95% relative maximum humidity |
Battery: | 1.5/1.55 V; SR41 |
Battery lifetime: | approx. 2700 measurements (using a new battery) |
IP Class: | IP22 |
Reference to standards: | EN 12470-3, clinical thermometers; ASTM E1112; IEC 60601-1; IEC 60601-1-2 (EMC); IEC 60601-1-11 |
Expected service life: | 5 years or 10000 measurements |
This device complies with the requirements of the Medical Device Directive 93/42/EEC.
echnical alterations reserved.
We grant you a 5 year guarantee after the date of purchase. Any damage caused by improper handling shall not be covered by the guarantee. The battery and packaging are excluded from the guarantee. All other damage claims excluded. A guarantee claim must be submitted with the purchase receipt. Please pack your defective instrument well and send with sufficient postage to the Microlife distributor.
§ Greek national patent No: 1007847 (201101007847)/31.10.2011;
§ International Patent Application No: WO/2013/064847;
§ European Patent Application No: 12798356.7/17.10.2012.
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Microlife MT 410 - Antimicrobial Digital Thermometer Manual