 Using a new probe cover each time, ensures this thermometer
is completely hygienic for use by the whole family.
Fever alarm
10 short beeps and a red LCD backlight alert the patient that he/
she may have a temperature equal to or higher than 37.5 °C.
3. Important Safety Instructions
 Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
 Never immerse this device in water or other liquids. For
cleaning please follow the instructions in the «Cleaning
and Disinfecting» section.
 Do not use this device if you think it is damaged or notice
anything unusual.
 Never open this device.
 Earwax in the ear canal may cause a lower temperature reading.
Therefore it is important to ensure the subject's ear canal is clean.
 Only use this thermometer with a new Microlife branded and
undamaged probe cover to prevent cross-infection.
 If the measurement result is not consistent with the patient's
finding or unusually low, repeat the measurement every 15
minutes or double check the result by another core body
temperature measurement.
 This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section.
 Ensure that children do not use this device unsupervised; some
parts are small enough to be swallowed.
 Do not use this device close to strong electromagnetic fields such
as mobile telephones or radio installations. Keep a minimum
distance of 3.3 m from such devices when using this device.
 Protect it from:
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
3
 If the device is not going to be used for a prolonged period the
battery should be removed.
WARNING: The measurement results given by this device
is not a diagnosis. It is not replacing the need for the consul-
tation of a physician, especially if not matching the patient's
symptoms. Do not rely on the measurement result only,
always consider other potentially occurring symptoms and
the patient's feedback. Calling a doctor or an ambulance is
advised if needed.
4. How this Thermometer measures Temperature
This thermometer measures infrared energy radiated from the
eardrum and the surrounding tissue. This energy is collected
through the lens and converted to a temperature value. The
measured reading obtained directly from the eardrum (Tympanic
Membrane) ensures the most accurate ear temperature.
Measurements taken from the surrounding tissue of the ear canal
generate lower readings and may result in misdiagnosis of fever.
To avoid an inaccurate measurement
1. Fit a new Microlife branded and undamaged probe cover CK on
the measuring sensor 1.
2. Switch on the thermometer by pressing the START/IO button 6.
3. After one beep is heard (and the temperature scale icon is
flashing), straighten the ear canal by gently pulling the middle
of the ear back and up.
4. Place the probe 1 firmly into the ear canal. «Good» will be
displayed with a short beep to confirm the device has detected
the correct position. Press the START/IO button 6 and keep
the probe in the ear until the thermometer beeps to signal the
completion of the measurement.
5. Control Displays and Symbols
 All segments displayed 7: Press the START/IO button 6 to
turn on the unit; all segments will be shown for 1 second.
 Ready for measurement 9: When the unit is ready for
measurement, the «°C» or «°F» icon will keep flashing.
The probe LED light is activated and will keep flashing.
 Correct position indication AT: The probe LED light will stop
flashing (stays illuminated) and «good» will be displayed on the
LCD, when the measuring sensor detects an appropriate position.
 Measurement complete AK: The reading will be shown on the
display 2 with the «°C» or «°F» icon; the unit is ready for the
next measurement, when the «°C» or «°F» icon is flashing again.