Advertisement
The measurement result given by this device is not a diagnosis! Do not rely on the measurement result only. | |
Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations, not with domestic waste. | |
Read the instructions carefully before using this device. | |
Type BF applied part | |
Keep dry |
To turn on the thermometer, press the ON/OFF button ; a short beep signals «thermometer ON». A display test is performed. All segments should be displayed.
Then at an ambient temperature of less than 32°C, an «L» and a flashing «°C» appear at the display field . The thermometer is now ready for use.
Correct functioning of the thermometer is tested automatically each time it is turned on. If a malfunction is detected (measurement inaccuracy), this is indicated by «ERR» on the display, and a measurement becomes impossible. In this case, the thermometer must be replaced.
Choose the preferred measuring method. When taking a measurement, the current temperature is continuously displayed and the «°C» symbol flashes. If the beep is heard 10 times and the «°C» is no longer flashing, this means that the measured increase in temperature is less than 0.1°C in 16 seconds and that the thermometer can now be read. To prolong the battery life, turn off the thermometer by briefly pressing the ON/OFF button . Otherwise the thermometer will automatically turn off after about 10 minutes.
If the ON/OFF button is pressed for more than 3 seconds when turning on the thermometer, the automatically stored maximum temperature during the last measurement will be displayed. At the same time, a «M» for memory will appear on the display. About 2 seconds after the button is released, the temperature value disappears and the thermometer is ready for measurement.
Use an alcohol swab or cotton tissue moistened with alcohol (70% Isopropyl) to clean the thermometer. Avoid wiping the display with Isopropyl alcohol to protect it from fading.
Please never immerse into liquids!
When the «» symbol (upside-down triangle) appears at the display field, the battery is flat and needs replacing. To replace the battery remove the battery compartment cover from the thermometer. Insert the new battery with the + at the top. Make sure you have a battery of the same type to hand. Batteries can be purchased at any electrical store.
Type: | Maximum thermometer |
Measurement range: | 32.0°C to 43.9°C Temp. < 32.0°C: display «L» for low (too low) Temp. > 43.9°C: display «H» for high (too high) |
Measurement accuracy: | ± 0.1°C between 34°C and 42°C |
Operating conditions: | 10 - 40°C; 15-95% relative maximum humidity |
Storage conditions: | -25 - +60°C; 15-95% relative maximum humidity |
Battery: | 1.5/1.55 V; LR41 |
Battery lifetime: | approx. 4500 measurements (using a new battery) |
IP Class: | IP21 |
Reference to standards: | EN 12470-3, clinical thermometers; ASTM E1112; IEC 60601-1; IEC 60601-1-2 (EMC); IEC 60601-1-11 |
Expected service life: | 5 years or 10000 measurements |
This device complies with the requirements of the Medical Device Directive 93/42/EEC.
Technical alterations reserved.
We grant you a lifetime guarantee after the date of purchase. Any damage caused by improper handling shall not be covered by the guarantee. The battery and packaging are excluded from the guarantee. All other damage claims excluded. A guarantee claim must be submitted with the purchase receipt. Please pack your defective instrument well and send with sufficient postage to the Microlife distributor.
Microlife AG, Espenstrasse 139
9443 Widnau / Switzerland
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.