64/88
Unplug the lamps.
Clean the reflector.
Only utilize original lamps.
Only touch new lamps in the handling zone.
Test the safety switch to ensure that it works.
The safety switch ensures that the device switches off in the event of a
faulty filter panel or when a filter panel is missing.
Press the button.
If the button springs back automatically then it is functional.
Replace the filter panel.
12.16.5

Cleaning and disinfection of the sensor

Not applicable for this device.