Environmental conditions
Optimum ambient temperature:
between 59 °F (15 °C) and 89.6°F (32°C)
Storage temperature: between 5 °F (-15 °C) and 140 °F (+60 °C)
Air pressure: between 794 hPa and 1060 hPa
Relative humidity: between 30 % and 70 %
Please contact the customer service department before putting the device into
operation - see page 2.
3.8.3

Commissioning

DANGER!
Damaged mains power cable!
Danger to persons through electric shock or burns.
Only the manufacturer or specialist companies are
permitted to replace the mains power cable.
Initial commissioning will be undertaken by the manufacturer's own service
personnel or by a (Level 3) JK Factory Certified Technician. The device will
be handed over ready for use.
As the studio operator you are responsible for abiding by the electrical regula-
tions applicable at the installation site.
The unit incorporates special preventive measures with regard to the electro-
magnetic compatibility and must be installed and commissioned according to
the EMC instructions obtained from the assembly instructions.
If the device has not been operated or serviced for a period of one year or
longer, it must first be inspected by a (Level 3) JK Factory Certified Techni-
cian before being put back into operation. Failure to do so may result in per-
sonal injury or property damage.