Summary of Contents for PARAMOUNT BED 7B01001000A3
Page 1
Serious incident means any incident that directly or indirectly resulted in, may have resulted in, or may result in any of the following. • Death of a patient, user, or other person • Serious temporary or permanent adverse effects on the health of a patient, user, or other person • Serious public health threat PARAMOUNT BED CO., LTD. [Revised: November 2021]...
CONTENTS BEFORE USING THE BED …………………………………………………………… 1. About this bed …………………………………………………………………………………… 4 • Intended use and benefits …………………………………………………………………… 4 • Part names …………………………………………………………………………………… 4 2. Safety precautions ……………………………………………………………………………… 9 3. Product symbol definitions …………………………………………………………………… 26 • Bed …………………………………………………………………………………………… 26 • Side-rail integrated panel / Patient control panel ………………………………………… 27 •...
Page 3
2. Operation procedure …………………………………………………………………………… 60 • Using the side-rail integrated panel (excluding PA-H 612*, PA-H 614*) ……………… 60 • Using the patient control panel (excluding PA-H 612*, PA-H 614*) …………………… 66 • Using the handheld controller (PA-H 612*, PA-H 614* only) …………………………… 69 3.
Page 4
Values described for the operation range, dimensions, angles, and weights are subject to normal manufacturing tolerances. Except for special cases, words such as approximately and about are omitted in this instruction manual to improve readability. n For any queries, contact Paramount Bed or your distributor.
1 BEFORE USING THE BED 1. About this bed Intended use and benefits [Intended use] This bed is designed to help patients maintain a comfortable posture at medical institutions. Use the bed according to the patient’s medical condition. [Clinical benefit] •...
Page 6
1 BEFORE USING THE BED Numbers with this symbol mark indicate pages where a related explanation is provided. * This illustration uses the PA-H 621* for explanation. P. 66 Knee section Patient control panel Leg section Hip section Head section P.
Page 7
Numbers with this symbol mark indicate pages where a related explanation is provided. * This illustration uses the PA-H 612* for explanation. P. 29 P. 30 and 31 P. 30 and 31 P. 29 Footboard Head-end side-rail Foot-end side-rail Headboard Foot end P.
Page 8
1 BEFORE USING THE BED Numbers with this symbol mark indicate pages where a related explanation is provided. * This illustration uses the PA-H 612* for explanation. Knee section Leg section Hip section Head section P. 38 Accessory attachment hole P.
Page 9
Head end Cable storage hooks Applied parts * Parts of the bed that in normal use necessarily come into physical contact with the patient for the bed to perform its function Head end Foot end...
Foot-end side-rail, and at the top of the leg section. Do not remove or deface them. n If a safety label is removed or defaced, replace it with a new one obtained from Paramount Bed (see p. 118) or your distributor.
Page 12
1 BEFORE USING THE BED WARNING <Gap related notes> g Watch the gaps. n There are gaps in the bed, side-rail, etc. There is a risk of injury if part of the body (especially the head and neck) gets caught in those gaps. n Be especially careful to check a patient who might act in unexpected ways or cannot maintain their posture independently.
Page 13
WARNING g Do not operate the bed with part of the body placed in a gap. n A body part could get caught in a gap, resulting in injury. n Be especially careful to check a patient who might act in unexpected ways or cannot maintain their posture independently.
Page 14
1 BEFORE USING THE BED WARNING <Falling off> g Be careful to prevent the patient from falling off the bed even when side- rails are raised. n The patient may fall through a gap between side-rails, or between a board and side-rail, resulting in injury. n The patient may bend forward over a side-rail and fall off the bed, resulting in injury. n When a thick mattress is used, the effective height of the side-rails becomes lower.
Page 15
WARNING <Operating the bed> g Do not allow the bed to be operated by a person deemed incapable of un- derstanding bed operation (a child aged 12 or under or a person with de- mentia). n If a person deemed incapable of understanding the operation operates the side-rail integrated panel, patient control panel, handheld controller (standard or accessory), or BED Navi (accessory) by oneself, the body may get caught in the bed by mistake, resulting in unexpected injury.
Page 16
1 BEFORE USING THE BED WARNING g Do not perform bed operations such as raising the head or knee section with a restraint belt attached to the mattress deck. n The patient’s body may be squeezed resulting in injury or the bed may be damaged or deformed. n If the patient or attendant operates the side-rail integrated panel, patient control panel, handheld controller (standard or accessory), or BED Navi (accessory), the body may get caught in the bed by mistake, resulting in unexpected injury.
Page 17
WARNING <Using the bed> g Use the bed with the boards attached, except when providing treatment. Also, check the patient’s condition when detaching the board. n A body part may get caught in a gap between the board and other objects, such as a suitable product or peripheral device, resulting in injury.
Page 18
1 BEFORE USING THE BED WARNING g Do not attach the restraint belt to any location other than the restraint holder. The restraint belt may unexpectedly come off resulting in injury, or the bed may be damaged or deformed. g The bed should be transferred by at least two persons. n When transferring the bed, blind spots may occur in the direction of travel and the bed may collide with a person, resulting in injury.
Page 19
WARNING <Transferring and moving the bed> g When transferring the bed, watch your step. Your leg may bump into the side pedal or your foot may get caught by a caster, resulting in injury. g Lock casters at all times except when moving the bed. n When the patient gets into or out of the bed, the bed may move, resulting in unexpected injury.
Page 20
1 BEFORE USING THE BED WARNING g Do not allow dust to accumulate on the power plug. n Dust on the power plug can absorb moisture and conduct electricity, resulting in a short circuit, electric shock, or fire. n Use a dry cloth to wipe away any dust that accumulates on the surface of the power plug.
Page 21
WARNING g Take care when handling electrical components such as the side-rail inte- grated panel or BED Navi (accessory). n The bed may malfunction resulting in injury or damage. In addition, a short circuit may cause electric shock or fire. •...
Page 22
Request inspection or repair of a damaged bed. If the bed is damaged in an earthquake, fire, flood, or any other incident, contact Paramount Bed (see p. 118) or your distributor for inspection or repair. Otherwise, there is a risk of electric shock or fire due to a short circuit or current leakage from electrical components, or injury due to malfunction caused by deformation of the bed.
Page 23
WARNING g Do not disassemble the battery. g Do not throw the battery into fire. g Keep the battery away from fire. g Do not charge the battery with a non-suitable bed. g Do not use the battery with equipment other than suitable beds. g Do not apply a strong impact to the battery. g Do not recharge the battery or operate the bed with the bed upside down.
Page 24
1 BEFORE USING THE BED CAUTION g When adjusting the height or inclination angle, make sure that the bed does not hit a wall, beam, or power outlet on the wall. The bed, wall, or floor may be damaged or deformed.
Page 25
If a physician, nurse, or caregiver often gets on the bed, contact Paramount Bed (see p. 118) or your distributor for regular inspection of the bed. g Do not apply a load exceeding the safe working load on the bed.
Page 26
1 BEFORE USING THE BED CAUTION g Do not move the bed while holding the side-rails. This applies excessive force to the side-rails, resulting in damage or deformation of the side-rails. g Avoid moving the bed over uneven floors. This may cause damage to the bed, or deformation or failure of the bed.
Make sure that a mattress or other Refer to the instruction manual products used with the bed are suitable products designated by Paramount Bed. Refer to the instruction manual Locking steering (straight forward movement) KOHKIN SIAA mark...
1 BEFORE USING THE BED Side-rail integrated panel / Patient control panel Symbol Description Symbol Description Cardiac Chair motion up/down Height up/down Electric CPR Head and knee coordinated up/ down Nurse call Disable operation of the patient control panel and handheld Trendelenburg/reverse Trende- controller (accessory) lenburg...
Handheld controller (RC-1850E: accessory) Symbol Description Symbol Description Indicates products covered by Head and knee coordinated up/ the EU directive on waste down electrical and electronic equip- ment (WEEE). This is a 2D Code (GS1 Data- Head up/down Matrix) that encodes the UDI (Unique Device Identifier) required by the Unique Device Knee up/down...
1 BEFORE USING THE BED 4. Preparing the bed Attaching and detaching the headboard or footboard The headboard and footboard can be attached and detached. When washing the patient’s hair, detach the headboard. When performing lower limb training or giving a foot bath, detach the footboard. <Attaching>...
Side-rail n There are a total of 4 side-rails, 2 on the left and right, respectively. n side-rails prevent the patient from falling off the bed. Foot end Foot end Head end Head end Raised state Lowered state n Be careful to prevent a body part (especially the head or neck) from entering a gap between the side-rail and mattress deck or mattress.
Page 32
1 BEFORE USING THE BED <Raising the side-rail> Grip Raise the side-rail and push in until it clicks into place. Foot end Pull the side-rail toward you to make sure that the side-rail is locked. Head end <Lowering the side-rail> Use one hand to hold the grip of the side-rail and push it to the back (toward the center of the bed).
Extension frame The bed can be extended for tall patients by pulling out the extension frame and adding an extension mattress (accessory). When using the extension frame, use a suitable extension mattress WARNING (accessory: see p. 83). A body part may get caught in a gap between the bed and extension frame or side-rail, resulting in injury.
Page 34
1 BEFORE USING THE BED <Storing the extension frame> Detach the extension mattress (accessory). Pull the frame stoppers on both sides of the bed toward you and turn them either clockwise or counterclockwise to unlock. (See p. 32) Hold the extension frame and push the extension frame back a little. Turn the frame stoppers to the locked position and then set them to the slide state (half-engaged state).
(deterioration may be accelerated in a high temperature and hu- midity environment). If the performance of the stopper deteriorates or wheels are discolored or cracked, contact Paramount Bed (see p. 118) or your distributor for after-sales services, such as repair.
1 BEFORE USING THE BED Raising the foot end You can change the angle of the leg section so that the foot end is raised. [Foot end lowered state] <Switching method> Perform the knee up operation to raise the knee section.
Accessory rail Accessory rails are in the locations shown below on the right and left sides of the bed. These rails can be used to hang accessories such as drainage bags. n Two each on the left and right sides of the hip section (4 locations in total) Do not apply a load of 3 kg or more to an accessory rail.
1 BEFORE USING THE BED Restraint holder When fastening restraint belts, fasten them in the locations in the figure below. The restraint holder can also be used as an accessory holder. Do not operate the bed with restraint belts fastened. The restraint belts WARNING may apply pressure to the patient, resulting in injury.
Accessory attachment hole Accessory attachment holes are located at the head end and foot end of the bed in a total of 4 locations. They can be used to attach suitable products such as an IV pole and oxygen cylinder holder. Accessory attachment hole Head end Foot end...
1 BEFORE USING THE BED Roller bumper Roller bumpers are attached to the 4 corners of the bed. When the bed is moved, they can absorb any impact that occurs when the bed hits a wall or equipment. Head end Foot end Roller bumper Roller bumper...
Angle indicator A head of bed angle indicator is located on the outside of the Head-end side-rail and an inclination angle indicator (excluding PA-H 611*, PA-H 612*, PA-H 613*, PA-H 614*) is located on the outside of the Foot-end side-rail. The head of bed angle and inclination angle can be checked respectively. * The indicated angle values should be used as approximations.
1 BEFORE USING THE BED Connecting to the power supply Do not connect multiple cables to a single outlet. If other electrical equipment is plugged into the same outlet and the rated capacity of the outlet or WARNING extension cable is exceeded, the power cable and plug may overheat and cause fire.
Battery (excluding PA-H 611*, PA-H 612*, PA-H 613*, PA-H 614*) * For the PA-H 611*, PA-H 612*, PA-H 613*, PA-H 614*, a battery is available as a suitable product. <Part names> Cover The battery is located under the main frame at the Head end head end of the bed.
Page 44
1 BEFORE USING THE BED <Battery-powered bed operation> When the battery power is on, the bed can be operated by power from the battery while the power plug is not inserted into the outlet (or in the event of a power failure). Do not operate the bed electrically during the emergency head-down...
Page 45
<Checking the battery status> g The battery status is displayed as shown in the table below. Power lamp the BED Navi Battery status Battery status (accessory) / indicator lamp on indication on the Power status handheld Battery status the side-rail BED Navi controller integrated panel...
Page 46
1 BEFORE USING THE BED <Recharging the battery> Inserting the power plug of the bed into the outlet Power plug starts recharging of the battery, and recharging will complete automatically after 24 hours. The battery can be recharged regardless of whether the power is on or off.
Attaching and detaching the handheld controller (RC-1850E: accessory) (excluding PA-H 612*, PA-H 614*) A handheld controller can be attached to this bed. * For details on the handheld controller, see p. 73 to 77. n Select a suitable combination of the bed and handheld controller. If the combination is not suitable, the bed could move unexpectedly, resulting in injury or bed failure.
1 BEFORE USING THE BED Handheld controller hook The handheld controller of the type with a handheld controller and the optional handheld controller (RC-1850E) have a hook. When not using the handheld controller, hook it on the top of the Head-end side- rail.
For details on the suitable bed communication cable and suitable nurse call system, con- tact Paramount Bed (see p. 118) or your distributor. Connecting to the nurse call system enables a nurse call to be made at the following times.
2 FUNCTIONS AND OPERATIONS 1. Bed functions n When performing a positioning operation, make sure that the patient’s head, arms, legs, or other body parts have not entered the gap between the bed and side-rail, or between the head section and frame. They may get caught, resulting in injury.
Page 51
<Height adjustment> n You can adjust the bed height. You can adjust the height from the floor to Head end Foot end mattress deck top in the range shown below. For the dimensions, see the table below. 125 mm single-wheel with resin cover 35.8 to 80.8 cm 125 mm single-wheel with iron cover (total lock) 36.3 to 81.3 cm...
2 FUNCTIONS AND OPERATIONS Trendelenburg/reverse Trendelenburg (excluding PA-H 611*, PA-H 612*, PA-H 613*, PA-H 614*) <Trendelenburg> n You can lower the head end of the bed to a maximum angle of 15˚. (See p. 63) * When inclining the bed in a range from the Down Head end Foot end lowest position (see the table below) to less...
Cardiac Chair motion (excluding PA-H 611*, PA-H 612*, PA-H 613*, PA-H 614*) Cardiac Chair motion (“Cardiac Chair”) is a function to reduce the feeling of pressure or slippage to help the patient maintain a better posture by alternately or simultaneously performing the head up, knee up, and inclining operations.
The bed communication cable and relay unit may not be suitable depending on the type of nurse call system. For details, contact Paramount Bed (see p. 118) or your distributor. n To use BED EXIT ALARM, you need to set the notification function settings (for the setting procedure, see the instruction manual for the relay unit).
Page 55
WARNING g Notification is not made during bed operation (e.g., head up, height ad- justment) even if the patient gets out of the bed or sits up on the bed. Fur- thermore, getting into or out of the bed during bed operation may cause the patient to fall off or fall over, resulting in injury. If you consider that the patient may get in or out of the bed during bed operation, set Overall lock- out or Patient lock-out (see p.
Page 56
2 FUNCTIONS AND OPERATIONS WARNING g If the patient does not stay in bed without moving for 5 seconds or more after returning to the bed, detection may not work, notification may not be made, and the patient may fall off or fall over, resulting in injury. g If the patient does not stay in bed without moving for 5 seconds or more after bed operation, detection may not work, notification may not be made, and the patient may fall off or fall over, resulting in injury.
Page 57
<SIT-UP> When the patient sits up in bed, notification is made through the nurse call system. The setting period can be selected from 0 second, 1 second, and 3 seconds. * If the patient sits down on the edge of the bed or gets out of the bed within 4 minutes after sitting up, notification is made only when the patient sits up to avoid repeated notifications.
2 FUNCTIONS AND OPERATIONS X-ray cassette imaging (type with X-ray cassette tray only) n Disconnect the power plug from the outlet and turn the battery off when using X-ray equipment. The bed may malfunction resulting in injury or damage. WARNING n Do not perform X-ray imaging using a cassette tray while an air mattress or Nemuri SCAN is installed.
Page 59
<Cassette setting procedure> Set a cassette while the head section is in a horizontal state. n When unlocking the locking lever, pull it toward you all the way until it stops. Otherwise, the lever may be damaged or deformed when the cassette tray is operated.
Page 60
2 FUNCTIONS AND OPERATIONS n Operate the cassette tray slowly. Not doing so may cause the cassette to be out of position resulting in imaging problems. n Hold the grip when operating the cassette tray. Not doing so may cause your hand or fingers to be caught and injured.
2. Operation procedure Using the side-rail integrated panel (excluding PA-H 612*, PA-H 614*) The side-rail integrated panel is intended for operation by physicians, nurses, and other medical professionals. It is located on the outside of the left or right Head-end side-rail. n Do not apply excessive force to the side-rail integrated panel or touch it with sharp objects.
Page 62
2 FUNCTIONS AND OPERATIONS g Bed operation buttons 1 Head up/down buttons ...........Adjust the angle of the head section. 2 Knee up/down buttons ...........Adjust the angle of the knee section. 3 Height up/down buttons ..........Adjust the bed height. 4 Trendelenburg/reverse Trendelenburg buttons .....Adjust the inclination angle of the bed. (Excluding PA-H 611*, PA-H 612*, PA-H 613*, PA-H 614*) 5 Head and knee coordination buttons ......Adjust the angle of the head/knee sections.
Page 63
g Function buttons 1 Nurse call auxiliary button (PA-H 631*, PA-H 633* only (suitable product)) ..Used to make a nurse call. The bed and nurse call wall plate need to be connected with the bed communication cable (commercially available product). 2 Patient lock-out button ..
Page 64
2 FUNCTIONS AND OPERATIONS <Operation procedure> Head up/down button 1 to raise the head section and press the Press the button to lower it. * Releasing the button stops the operation at its position. Knee up/down button 2 to raise the knee section and press the Press the button to lower it.
Page 65
For details on the suitable nurse call system and suitable bed communication cable, con- tact Paramount Bed (see p. 118) or your distributor.
Page 66
2 FUNCTIONS AND OPERATIONS LED lamp Electric CPR button 8 for at least 2 seconds or press it twice quickly to operate the bed so the head/ Hold down the knee sections and inclination angles become 0˚ (horizontal). * The bed is lowered to the pause position. * Releasing the button stops the operation at its position.
Using the patient control panel (excluding PA-H 612*, PA-H 614*) <Part names and functions> The patient control panel is intended for patients to perform the following operations. It is located on the inside of the left or right Head-end side-rail. n Adjusting the angle of the head/knee sections n Charging USB devices (on the right side only when the patient is lying down) n Do not apply excessive force to the patient control panel or touch it with...
Page 68
For details on the suitable nurse call system and suitable bed communication cable, con- tact Paramount Bed (see p. 118) or your distributor. Head up/down button 1 on p. 66 to raise the head section and press the Press the button to lower it.
Page 69
<Using the USB port> USB port case Pull down the claw of the USB cover on the back of the USB port case to open the USB cover. You can connect a USB device to the port. After disconnecting the USB device, perform the procedure in 1 in reverse to insert the USB cover into the back of the USB port case.
2 FUNCTIONS AND OPERATIONS Using the handheld controller (PA-H 612*, PA-H 614* only) You can adjust the angle of the head and knee sections and the bed height with no increments with the handheld controller. The handheld controller is intended to be operated by patients. Do not allow it to be operated by a person deemed incapable of understanding the operation (a child aged 12 or under or a person with WARNING...
3. Emergency operation In an emergency, you can easily perform the cardiopulmonary resuscitation by operating the CPR lever attached to the head section or the electric CPR on the side-rail integrated panel. n When pressing the CPR button or operating the CPR lever, make sure that the patient’s, caregiver’s, and other people’s head, arms, legs, or other body parts have not entered the gap between the mattress deck and main frame.
• Regularly check that the wiring work has been performed properly. Suitable products may be changed due to specification changes or combinations. In addition, some suitable products may not be listed in the table due to new release or discontinuation of sales. For any queries on suitable products, contact Paramount Bed (see p. 118) or your distributor. Suitable product* Model No.*...
Use this bed in combination with a suitable mattress designated by Paramount Bed. If you use the bed in a combination with a mattress other than suitable mattresses designated by Paramount Bed, they may...
3 SUITABLE PRODUCTS 3. Accessories Handheld controller <Part names and functions> You can adjust the angle of the head and knee sections and the bed height with no increments with the handheld controller. The handheld controller is intended to be operated by patients.
Page 75
3 Wrench mark ..When it is time to replace the handheld controller, the wrench mark is displayed in the setting mode. Contact Paramount Bed (see p. 118) or your distributor. g Identification embossing can be identified by touching the embossed operation buttons.
Page 76
3 SUITABLE PRODUCTS <Setting switching> n Do not allow it to be operated by a person deemed incapable of understanding the operation (a child aged 12 or under or a person with dementia). If a person deemed incapable of understanding the operation operates the handheld controller by oneself, the body may get caught in the bed by mistake, resulting in accidental injury.
Page 77
g Setting each operation of the bed You can change the settings for BED SPEED and SOUND. n Setting BED SPEED of the head n Setting the operation sound and knee sections You can unmute or mute the You can switch BED SPEED of the operation sound that is emitted by head and knee coordination and pressing the operation button of the...
Page 78
3 SUITABLE PRODUCTS g Setting procedure n You cannot mute the alarm sound in the following cases. • When the height from floor to mat- tress deck top has reached 43 cm and the operation is stopped when Press the settings lowering the bed height.
*2) Nemuri SCAN is not designed for sales in Europe. Also, it is not distributed in some regions as well where it is not registered as medical equipment. For details on the sales of Nemuri SCAN, contact Paramount Bed (see p. 118) or your distributor. Model No.
* IV pole (IC-56A, IC-56B, IC-56C) is not designed for sales in Europe. Also, it is not distributed in some regions as well where it is not registered as medical equipment. For details on the sales of IV pole, contact Paramount Bed (see p. 118) or your distributor. Model No.
* Oxygen cylinder holder (IC-A83) is not designed for sales in Europe. Also, it is not distributed in some regions as well where it is not registered as medical equipment. For details on the sales of oxygen cylinder holder, contact Paramount Bed (see p. 118) or your distributor. <PC-5116/IC-A83>...
Page 82
3 SUITABLE PRODUCTS <IC-A84> Accessory attachment holes are located at the head end of the bed in a total of 2 locations. Head end Accessory attachment hole Oxygen cylinder holder (IC-A84) WARNING Use the oxygen cylinder holder (IC-A84) at position A in the figure on the left. If used in position B or C, the bed may become unstable and tip over, resulting in injury.
Also, it is not distributed in some regions as well where it is not registered as medical equipment. For details on the sales of traction frame, contact Paramount Bed (see p. 118) or your distributor. * This illustration uses the KA-83U model for Model No.
* Extension mattress (IE-52EB, IE-52ED) is not designed for sales in Europe. Also, it is not distributed in some regions as well where it is not registered as medical equipment. For details on the sales of extension mattress, contact Paramount Bed (see p. 118) or your distributor. Model No.
* Foot spacer (IC-A33321Q) is not designed for sales in Europe. Also, it is not distributed in some regions as well where it is not registered as medical equipment. For details on the sales of side-rail spacer/Foot spacer, contact Paramount Bed (see p. 118) or your distributor.
3 SUITABLE PRODUCTS Side-rail spacer holder Attach the side-rail spacer holder in the correct orientation. If attached in CAUTION the wrong orientation, it may hit the patient, resulting in injury. You can store the side-rail spacer/Foot spacer. Hang the side-rail spacer holder on the footboard outer side. When the side-rail spacer holder is being used, the following accessories cannot be attached to the footboard.
Lifting pole Model No. PC-5111 n Load capacity is 76 kg. Do not hang an article exceeding the load capacity because it may damage the lifting pole, resulting in injury. WARNING n During transportation, do not hang on to the lifting pole. It may tip over the bed, the patient on the bed may fall off the bed, resulting in injury.
3 SUITABLE PRODUCTS Monitor table n You can attach the monitor table to the footboard of the A6 Series and use it as a table to place a monitor. n Do not place an object exceeding 12 kg on the monitor table. The monitor table or bed may be damaged or deformed resulting in injury, or the placed object may be damaged or deformed.
Bed linen rack Model No. PC-5060 n Load capacity is 1 kg. Do not set an article exceeding the load capacity because it may cause the linen rack to be damaged or deformed. CAUTION n Attach the bed linen rack in the correct orientation. If attached in the wrong orientation, the rack may hit the patient, resulting in injury.
Also, it is not distributed in some regions as well where it is not registered as medical equipment. For details on the sales of chart holder, contact Paramount Bed (see p. 118) or your distributor. g The chart holder can be used for setting the charts.
*1) Nemuri SCAN is not designed for sales in Europe. Also, it is not distributed in some regions as well where it is not registered as medical equipment. For details on the sales of Nemuri SCAN, contact Paramount Bed (see p. 118) or your distributor. Model No.
3 SUITABLE PRODUCTS 4. Peripherals Bedside table The bedside table is used for having meals or doing easy chores on the bed. Bedside table Model No. KF-1900, KF-1920, KF-1930, KF-1950, KF-1960, KF-1970, KF-1924, KF-1934, KF-1954, Head end KF-1964, KF-1974, KF-282A, PF-3100A, PF-3110A, PF-3120A, PF-3130A Foot end n Do not perform bed operations when the bedside table is placed on the...
*1) Sittan™ and rehabilitation table are not sold in some countries or regions. Also, it may be classified as medical equipment. For details, contact Paramount Bed (see p. 118) or your distributor. Model No.
*1) The lift is not sold in some countries or regions. Also, it may be classified as medical equipment. For details, contact PARAMOUNT BED CO., LTD. (see p. 118) or your distributor. n Also read the instruction manual supplied with the lift. Improper use may result in accidental injury.
Problems may not always indicate faults. Check the following once again before requesting repair. g If the problem persists after taking corrective action, stop using the bed immediately, disconnect the power plug from the outlet, turn the battery off, and then contact Paramount Bed (see p. 118) or your distributor for repair.
Page 97
For details on the suitable nurse call system and suitable Are the nurse call system and bed communica- ― bed communication cable, contact Paramount Bed (see tion cable suitable for the bed? p. 118) or your distributor. Is the notification lamp on the relay unit lit? Turn on notification on the relay unit.
Page 98
4 IN CASE OF TROUBLE Reference Symptoms and status Check Corrective action page Are the notification settings configured properly? Configure the notification settings properly. If an object of 3 kg or more is removed from the bed, Is an object of 3 kg or more removed from the notification may be made. In such a case, turn off or bed? pause notification.
Page 99
g Positions of product ID labels Product ID labels are affixed to the right side of the main frame and on the inside of the footboard. Product ID label Head end Foot end Footboard Product ID label * Affixed to the underside of the board.
Inspect for loose parts and damage and the operation of the moving parts regularly. If you notice an irregularity or want a detailed inspection, contact Paramount Bed (see p. 118) or your distributor.
5 REFERENCE INFORMATION Daily maintenance n To prevent accidents caused by incorrect operation, always disconnect the power plug from the outlet before maintenance. WARNING n Do not clean the bed by pouring water onto it. Short circuit and electric shock may occur. <Bed>...
2. Long-term storage and disposal Long-term storage The following precautions should be observed when the bed will not be in use for an extended period of time. n Turn off BED EXIT ALARM notification. n Lay the head and knee sections flat. n Lower the bed height to the minimum height.
Transport Humidity 30 to 90% RH and storage Atmospheric conditions 620 to 1060 hPa pressure 8 years (based on Paramount Bed data) * On condition that you perform the specified maintenance and inspection Service life and replace the consumable parts (e.g., casters). Headboard/ HDPE resin molding footboard Bed side-rail HDPE resin molding...
Page 105
Product name Electrical Hospital Bed PA-H 6115 PA-H 6135 PA-H 6215 PA-H 6235 PA-H 6125 PA-H 6145 PA-H 6315 PA-H 6335 Model No.* PA-H 6115 X PA-H 6135 X PA-H 6215 X PA-H 6235 X PA-H 6125 X PA-H 6145 X PA-H 6315 X PA-H 6335 X Type...
Page 106
Transport Humidity 30 to 90% RH and storage Atmospheric conditions 620 to 1060 hPa pressure 8 years (based on Paramount Bed data) * On condition that you perform the specified maintenance and inspection and re- Service life place the consumable parts (e.g., casters). Headboard/ HDPE resin molding footboard Bed side-rail HDPE resin molding...
Page 107
Product name Electrical Hospital Bed PA-H 6117 PA-H 6137 PA-H 6217 PA-H 6237 PA-H 6127 PA-H 6147 PA-H 6317 PA-H 6337 Model No.* PA-H 6117 X PA-H 6137 X PA-H 6217 X PA-H 6237 X PA-H 6127 X PA-H 6147 X PA-H 6317 X PA-H 6337 X Type...
Page 108
Transport Humidity 30 to 90% RH and storage Atmospheric conditions 620 to 1060 hPa pressure 8 years (based on Paramount Bed data) * On condition that you perform the specified maintenance and inspection and re- Service life place the consumable parts (e.g., casters). Headboard/ HDPE resin molding footboard Bed side-rail HDPE resin molding...
Page 109
Product name Electrical Hospital Bed PA-H 6118 PA-H 6138 PA-H 6218 PA-H 6238 PA-H 6128 PA-H 6148 PA-H 6318 PA-H 6338 Model No.* PA-H 6118 X PA-H 6138 X PA-H 6218 X PA-H 6238 X PA-H 6128 X PA-H 6148 X PA-H 6318 X PA-H 6338 X Type...
Page 110
Transport Humidity 30 to 90% RH and storage Atmospheric conditions 620 to 1060 hPa pressure 8 years (based on Paramount Bed data) * On condition that you perform the specified maintenance and inspection and re- Service life place the consumable parts (e.g., casters). Headboard/ HDPE resin molding footboard Bed side-rail HDPE resin molding...
Page 111
Product name Electrical Hospital Bed PA-H 6119 PA-H 6139 PA-H 6219 PA-H 6239 PA-H 6129 PA-H 6149 PA-H 6319 PA-H 6339 Model No.* PA-H 6119 X PA-H 6139 X PA-H 6219 X PA-H 6239 X PA-H 6129 X PA-H 6149 X PA-H 6319 X PA-H 6339 X Type...
Page 112
Transport Humidity 30 to 90% RH and storage Atmospheric conditions 620 to 1060 hPa pressure 8 years (based on Paramount Bed data) * On condition that you perform the specified maintenance and inspection and re- Service life place the consumable parts (e.g., casters). Headboard/ HDPE resin molding footboard Bed side-rail HDPE resin molding...
Page 113
Product name Electrical Hospital Bed PA-H 6110 PA-H 6130 PA-H 6210 PA-H 6230 PA-H 6120 PA-H 6140 PA-H 6310 PA-H 6330 Model No.* PA-H 6110 X PA-H 6130 X PA-H 6210 X PA-H 6230 X PA-H 6120 X PA-H 6140 X PA-H 6310 X PA-H 6330 X Type...
Page 114
5 REFERENCE INFORMATION n Essential performance Function to support the body of patient, Emergency CPR (head down) function. n Classification and Standards The electric hospital bed is designed and manufactured according to the following equipment classification and standards. • Classification as medical device in Europe : Class 1 •...
4. Technical explanation regarding EMD (Electromagnetic Disturbance) The bed requires special safety precautions regarding EMC to be observed. Install, operate, and use the bed according to the information provided below. The bed is intended for use in an electromagnetic environment specified according to “Electromagnetic interference.”...
Page 116
5 REFERENCE INFORMATION Guidance and manufacturer’s declaration — electromagnetic immunity The bed is intended for use in the following electromagnetic environment. The customer or user of the bed must use it in an environment specified here. IEC 60601-1-2 Immunity test Compliance level Test level Contact discharge: ±8 kV Contact discharge: ±8 kV Electrostatic discharge (ESD)
Page 117
Guidance and manufacturer’s declaration — electromagnetic immunity The bed is intended for use in the following electromagnetic environment. The customer or user of the bed must use it in an environment specified here. IEC 60601-1-2 Electromagnetic environment Immunity test Compliance level Test level —...
Page 118
5 REFERENCE INFORMATION Guidance and manufacturer’s declaration — electromagnetic immunity The bed is intended to be used in an electromagnetic environment where RF radio communication devices are used. The electromagnetic interference between the RF communication device and the bed can be prevented by maintaining the minimum distance between the two as recommended below in accordance with the maximum output and frequency of the communication device.
“1. Troubleshooting” (see p. 95 to 98) in this instruction manual. If the problem persists after taking corrective action, disconnect the power plug of the bed from the outlet and turn the battery off, and then contact Paramount Bed (see Contact Information below) or your distributor. g Details necessary when requesting repair n Product name, model number, number on product ID label (see p.
Page 120
Instruction manual edited by Torindo Co., Ltd. 2021-11...