Attaching And Detaching The Handheld Controller (Rc-1850E: Accessory) (Excluding Pa-H 612*, Pa-H 614*) - PARAMOUNT BED 7B01001000A3 Operator's Manual

Electrical hospital bed
Table of Contents

Advertisement

Attaching and detaching the handheld controller (RC-1850E: accessory) (excluding PA-H 612*, PA-H 614*)

A handheld controller can be attached to this bed.
* For details on the handheld controller, see p. 73 to 77.
WARNING
There are handheld controller power outlets located one each on the left and right at the center of the bed
main frame.
Foot end
Handheld controller power outlet
Guide lock
Plug
n Select a suitable combination of the bed and handheld controller. If the
combination is not suitable, the bed could move unexpectedly, resulting in
injury or bed failure. (For the model numbers and shapes of the handheld
controller, see p. 73 to 77.)
n Disconnect the power plug of the bed from the outlet and turn the battery off
before attaching the handheld controller to the bed. If the handheld controller
is inserted into the bed while the bed is plugged into the outlet, the bed could
move unexpectedly, resulting in injury or bed failure.
Handheld controller power outlet
Insert or pull out the handheld controller
straight.
<Attaching>
1.
Insert the plug into the handheld controller power
outlet.
2.
To prevent the plug from being disconnected,
turn the guide lock clockwise to lock.
<Detaching>
1.
Turn the guide lock counterclockwise to unlock.
2.
Pull out the plug.