Connecting The Bed Communication Cable (Commercially Available Product) (Pa-H 631*, Pa-H 633* Only) - PARAMOUNT BED 7B01001000A3 Operator's Manual

Electrical hospital bed
Table of Contents

Advertisement

Connecting the bed communication cable (commercially available product) (PA-H 631*, PA-H 633* only)

WARNING
n Be sure to check the operation after connecting the bed communication cable.
n For details on the suitable bed communication cable and suitable nurse call system, con-
tact Paramount Bed (see p. 118) or your distributor.
Connecting to the nurse call system enables a nurse call to be made at the following times.
n When BED EXIT ALARM detects a change in the status of the patient (For details on BED EXIT ALARM,
see p. 53 to 56.)
n When the nurse call auxiliary button of the bed is pressed (For details on the nurse call auxiliary button,
see p. 62, 66, and 67.)
Bed side connector
Bed communication cable
(37-pin female)
n Use a suitable bed communication cable. A nurse call may not be made or an
equipment failure may occur when an unsuitable cable is used.
n Before moving the bed, disconnect the bed communication cable from the
bed or the nurse call system on the wall. Furthermore, do not apply
excessive force to the bed communication cable or run over it with the bed. A
nurse call may not be made due to an equipment or cable failure.
(37-pin male)
<Connecting the cable>
1.
Connect the bed communication cable (37-pin
female) to the bed side connector of the bed.
2.
Connect the other end of the bed communication
cable (37-pin male) to the nurse call wall plate.
<Checking the operation>
1.
Insert the power plug into the outlet.
2.
Press the nurse call auxiliary button located at
the Head-end side-rail.
3.
Check that the LED lamp above the nurse call
auxiliary button turns on and a nurse call is
made.
48
LED lamp
Nurse call auxiliary
button