5 REFERENCE INFORMATION

1. Regular inspection and daily maintenance

Regular inspection

To ensure safety, inspect the product regularly, at least once a year. The product wears out and deteriorates
differently depending on the operating frequency and environment. Inspect for loose parts and damage and
the operation of the moving parts regularly.
If you notice an irregularity or want a detailed inspection, contact Paramount Bed (see p. 118) or your
distributor.
Part
Height adjustment operation
Bed
Board
Caster
Caster locking operation
Side-rail
Visual check of electrical components
Electrical components
Nurse call notification check (only
when using the nurse call auxiliary
Operation
Proper operation can be achieved.
Head up operation
* See p. 49
There must not be any strange noises.
Proper operation can be achieved.
Knee up operation
* See p. 49
There must not be any strange noises.
Proper operation can be achieved.
* See p. 50
There must not be any strange noises or significant rattling.
Bed operation buttons
All bed operation buttons must work.
* See p. 60 to 67
CPR
The head section must be lowered with a CPR lever.
* See p. 70
There must not be any deformation, damage, or detachment of roller bumpers or
resin parts.
Miscellaneous
There must not be any peeling of labels, and the text must be legible.
There must not be any scratches or deformation at the mattress deck, frame, or
any other parts.
The board stopper must be locked properly.
Attachment state
* See p. 29
There must not be any significant rattling.
There must not be any strange noises when the bed is traveling.
Traveling
There must not be any abnormal vibrations when the bed is traveling.
Proper operation can be achieved.
The operation must not be overly difficult.
* See p. 34
The bed must not move when casters are locked.
The bed must move smoothly when the lock is released.
Proper operation can be achieved.
Side-rail operation
There must not be any strange noises or significant rattling.
* See p. 30 and 31
The operation must not be overly difficult.
Cables must not get tangled in the bed.
The plug and the cable must have no scratches or damage.
The side-rail integrated panel, patient control panel, and handheld controller (stan-
dard or accessory), BED Navi (accessory) must have no scratches or damage.
The pins of the bed side connector (37-pin male) must not be damaged.
The LED lamp above the nurse call auxiliary button must turn on and a nurse call
must be made when the nurse call auxiliary button is pressed while the bed is
button of the bed)
connected to the nurse call wall plate.
Inspection item