Bed Exit Alarm (Pa-H 631*, Pa-H 633* Only) - PARAMOUNT BED 7B01001000A3 Operator's Manual

Electrical hospital bed
Table of Contents

Advertisement

2 FUNCTIONS AND OPERATIONS

BED EXIT ALARM (PA-H 631*, PA-H 633* only)

g This function measures load changes on the bed with sensors built into the bed and notifies the
status of a patient who may fall off or fall down.
g The following notification functions can be used with BED EXIT ALARM. BED EXIT ALARM causes
an alert when it detects the patient's movement. Also a notification can be made to the nurse call
system by connecting it to the nurse call wall plate with a nurse call relay unit (accessory) ("relay
unit") or a bed communication cable.
n MONITORING
n LEFT BED
n EDGE OF BED
n SIT-UP
Each function is set using the relay unit. For connecting to the nurse call system and details on handling,
see "Connecting the bed communication cable" (see p. 48) or the instruction manual for the relay unit.
g If a power failure occurs, all bed functions including BED EXIT ALARM stop, so the patient status
cannot be notified by BED EXIT ALARM. If necessary, connect the power plug of the bed to an emer-
gency power supply.
n This product is just an assisting device to detect out-of-bed and does not guarantee safety.
A false notification or absence of notification may occur depending on the motion or lying
position of the patient. We assume no responsibility whatsoever for any damages resulting
from an accident, etc.
n The bed communication cable and relay unit may not be suitable depending on the type of
nurse call system. For details, contact Paramount Bed (see p. 118) or your distributor.
n To use BED EXIT ALARM, you need to set the notification function settings (for the setting
procedure, see the instruction manual for the relay unit).
n Before using BED EXIT ALARM for the first time after the bed is installed, make sure that
notification is made correctly before use.
n If notification is not made or any other errors occur, stop using BED EXIT ALARM and
check the bed and bed communication cable, or relay unit by referring to "1. Troubleshoot-
ing" on p. 95 to 98.
n If you turn the notification function on while the bed height is less than "+4 cm" from the
minimum height, the bed rises to the "minimum height +4 cm."
53

Advertisement

Table of Contents
loading

This manual is also suitable for:

A6 series

Table of Contents